Avtech product is found to be defective, Avtech shall at its option repair or replace said defective item. This warranty does not apply to units which have been dissembled, modified or subjected to conditions exceeding the applicable specifications or ratings.
TABLE OF CONTENTS WARRANTY........................2 TECHNICAL SUPPORT....................2 TABLE OF CONTENTS....................3 INTRODUCTION......................5 -KMPA OPTION, -PW1 OPTION....................6 SPECIFICATIONS......................7 EUROPEAN REGULATORY NOTES................8 EC DECLARATION OF CONFORMITY..................8 DIRECTIVE 2002/95/EC (RoHS)....................8 INSTALLATION....................... 9 VISUAL CHECK........................9 CONNECTING THE OUTPUT MODULE TO THE MAINFRAME..........9 POWER RATINGS........................9 CONNECTION TO THE POWER SUPPLY................9 PROTECTION FROM ELECTRIC SHOCK................10...
INTRODUCTION The Model AV-107C-B pulse generator is designed for pulsing laser diode and other low impedance loads with rectangular pulses as high as 10 Amperes into load voltages up to 60V, with 20 ns rise and fall times. The pulse repetition frequency can vary from 1 Hz to 5 kHz, and pulse widths can vary from 100 ns to 1 us (or optionally 20 us).
The AV-107C-B has two amplitude ranges, of 0 to 1A and 1 to 10A, allowing the instrument to be used at both moderate and high current levels. (The instrument automatically selects the appropriate range based on the amplitude setting.) The AV-107C-B can be controlled from the front panel, or via a computer connected to the IEEE 488.2-compliant GPIB port, or the RS-232 serial port.
SPECIFICATIONS Model: AV-107C-B GPIB, RS-232 control: Standard on -B units. Amplitude: -P units: 0 to +10 Amperes -N units: 0 to -10 Amperes -PN units: 0 to ±10 Amperes Pulse width: 100 ns to 1 us (with -KMPA or -PW1 option: 100 ns to 20 us) Rise time, fall time: ≤...
This instrument is exempt from Directive 2002/95/EC of the European Parliament and of the Council of 27 January 2003 on the Restriction of the use of certain Hazardous Substances (RoHS) in electrical and electronic equipment. Specifically, Avtech instruments are considered "Monitoring and control instruments" (Category 9) as defined in Annex 1A of Directive 2002/96/EC.
INSTALLATION VISUAL CHECK After unpacking the instrument, examine to ensure that it has not been damaged in shipment. Visually inspect all connectors, knobs, liquid crystal displays (LCDs), and the handles. Confirm that a power cord, a GPIB cable, and two instrumentation manuals (this manual and the “Programming Manual for -B Instruments”) are with the instrument.
5) Do not attempt any repairs on the instrument, beyond the fuse replacement procedures described in this manual. Contact Avtech technical support (see page 2 for contact information) if the instrument requires servicing. Service is to...
ENVIRONMENTAL CONDITIONS This instrument is intended for use under the following conditions: 1. indoor use; 2. altitude up to 2 000 m; 3. temperature 5 °C to 40 °C; 4. maximum relative humidity 80 % for temperatures up to 31 °C decreasing linearly to 50 % relative humidity at 40 °C;...
FUSES This instrument contains four fuses. All are accessible from the rear-panel. Two protect the AC prime power input, and two protect the internal DC power supplies. The locations of the fuses on the rear panel are shown in the figure below: Fuses #1 and #2 Fuse #4 Fuse #3...
DC FUSE REPLACEMENT The DC fuses may be replaced by inserting the tip of a flat-head screwdriver into the fuse holder slot, and rotating the slot counter-clockwise. The fuse and its carrier will then pop out. FUSE RATINGS The following table lists the required fuses: Nominal Manufacturer’s Distributor’s...
MAINFRAME FRONT PANEL CONTROLS 1. POWER Switch . This is the main power switch. When turning the instrument on, there may be a delay of several seconds before the instrument appears to respond. 2. OVERLOAD Indicator . When the instrument is powered, this indicator is normally green, indicating normal operation.
Page 15
5. KEYPAD . Control Name Function MOVE This moves the arrow pointer on the display. CHANGE This is used to enter the submenu, or to select the operating mode, pointed to by the arrow pointer. ×10 If one of the adjustable numeric parameters is displayed, this increases the setting by a factor of ten.
REAR PANEL CONTROLS GRN=LNK GATE YEL=ACT TRIG RS-232 GPIB 1. AC POWER INPUT . An IEC-320 C14 three-pronged recessed male socket is provided on the back panel for AC power connection to the instrument. One end of the detachable power cord that is supplied with the instrument plugs into this socket. 2.
Page 17
“Programming Manual for -B Instruments” for more details on RS-232 control. 8. LAN Connector and Indicator . (Optional feature. Present on -TNT units only.) The -TNT option "Internet-enables" Avtech pulse generators by adding this standard Ethernet port to the rear panel, in addition to the IEEE-488.2 GPIB and RS-232 ports normally found on "-B"...
If the load cannot be placed directly on the output terminals of the -PG module, the AV- LZ lines should be used between the -PG module and the load (see the Avtech AV-LZ data sheet, available at www.avtechpulse.com).
GENERAL INFORMATION BASIC TEST ARRANGEMENT Diode 5-foot 25-pin cable (device under test) OUTPUT AVTECH REAL-TIME MODULE MAINFRAME OSCILLOSCOPE OUT (25 PIN) CONTROL scope probe HV (BNC) CHANNEL A series resistance POWER 5-foot TRIG SYNC RG-58 INPUT OUTPUT coaxial cable The diode is shown oriented for positive (-P) operation.
Beware that wirewound resistors usually have far too much inductance to be useful as current-sensing resistors. 4. The AV-107C-B includes a monitor output connector on the output module. The monitor output should be terminated with a 50 Ohm load. On standard units, the...
Page 21
pulse width are variable. The SYNC pulse is a fixed-width TTL-level reference pulse used to trigger oscilloscopes or other measurement systems. When the delay is set to a positive value the SYNC pulse precedes the OUT pulse. These pulses are illustrated below, assuming internal triggering, positive delay, and a positive output: 100 ns, FIXED SYNC OUT...
> 50 ns TRIG TTL LEVELS (external input) (0V and 3V-5V) PROPAGATION DELAY (FIXED) 100 ns, FIXED SYNC OUT 3V, FIXED DELAY > 0 PULSE WIDTH AMPLITUDE, VARIABLE MAIN OUTPUT Figure C As before, if the delay is negative, the order of the SYNC and OUT pulses is reversed. The delay, pulse width, and frequency (when in the internal mode), of the OUT pulse can be varied with front panel controls or via the GPIB or RS-232 computer interfaces.
GATING MODES Triggering can be suppressed by a TTL-level signal on the rear-panel GATE connector. The instrument can be set to stop triggering when this input high or low, using the front- panel gate menu or the appropriate programming commands. This input can also be set to act synchronously or asynchronously.
LOAD PROTECTION This instrument is designed to pulse resistive and diode loads and will exhibit a large output spike when used to drive a load with significant inductance (as predicted by LENZ'S LAW). For this reason the load should be connected to the output using low inductance leads (as short as possible and as heavy a gauge as possible).
START-UP CHECK-LIST FOR LOCAL CONTROL The instruction manual has been studied thoroughly. The “Local Control” section of the “Programming Manual for -B Instruments” has been studied thoroughly. The output module is connected to the mainframe as shown in the “Basic Test Arrangement”...
Page 26
initially be pointing at the “Output Off” choice. Press MOVE so that the arrow pointer is pointing at the “Output On” choice. (The mainframe is now supplying a trigger to the output module.) Press CHANGE to return to the main menu. 12) Press the MOVE button to move the arrow pointer to the amplitude menu item.
CALIBRATION ADJUSTMENTS - SOFTWARE PROCEDURES ADJUSTING AMPLITUDE ACCURACY The AV-107C-B has two amplitude ranges: 0 to 1A, and 1A to 10A, approximately. The calibration of each range can be adjusted by a few percent if necessary. If it is found that the output amplitude settings (as set by the front-panel controls or programming commands) do not agree exactly with measured values of amplitude (i.e.,...
PROGRAMMING YOUR PULSE GENERATOR KEY PROGRAMMING COMMANDS The “Programming Manual for -B Instruments” describes in detail how to connect the pulse generator to your computer, and the programming commands themselves. A large number of commands are available; however, normally you will only need a few of these.
ALL PROGRAMMING COMMANDS For more advanced programmers, a complete list of the available commands is given below. These commands are described in detail in the “Programming Manual for -B Instruments”. (Note: this manual also includes some commands that are not implemented in this instrument.
There are no user-adjustable internal circuits. For repairs other than fuse replacement, please contact Avtech (info@avtechpulse.com) to arrange for the instrument to be returned to the factory for repair. Service is to be performed solely by qualified service personnel.
MAINTENANCE REGULAR MAINTENANCE This instrument does not require any regular maintenance. On occasion, one or more of the four rear-panel fuses may require replacement. All fuses can be accessed from the rear panel. See the “FUSES” section for details. CLEANING If desired, the interior of the instrument may be cleaned using compressed air to dislodge any accumulated dust.
Do not attempt any repairs on this ins trument Mount using four M2.5-0.45X12 screws. beyond the fuse replacement procedures described in the manual. Contact Avtech if the instr ument requires servicing. Service is to be perf or med WARN ING solely by qualified se rvice per sonnel.
PCB 158H - LOW VOLTAGE DC POWER SUPPLY, 1/3 p cb 1 58 h _ ov p p cb 1 58 h _ ov p .sch +1 5V +1 5V +1 5V GN D GN D P-OU T # 1 P-OU T # 1 -1 5 V -1 5 V...
PCB 158H - LOW VOLTAGE DC POWER SUPPLY, 2/3 8 3 08 3 5 FUSEHO LDER 6 4 04 4 5 -6 TEST-LOO P TEST-LOO P S1 A S1 B, OR D C S2 A, O R DC S2 B 4 3 4- 13 -1 0 0 M C2 0 C1 9...
PCB 158H - LOW VOLTAGE DC POWER SUPPLY, 3/3 BLEED ER S D1 1 R2 1 CAPBANK 3 3 0 OY 1 .5 K O Y 1 N4 9 3 7 C2 3 1 0 00 u F,3 5 V (P5 16 9 -ND ) SHO RTS OU T B ASE WHEN C HAR GING.
PCB 160B - HIGH VOLTAGE DC POWER SUPPLY SMA 2 1 -6 4 04 5 6 -0 SMA 1 HV WA RN ING RELA Y, G 2R-1 -E, 2 4 V TR25 0 -0 8 0U 2 0 0 2 0 0 G6 A -2 -H,12 V 1 N4 9 3 7 CHASSIS GND...
PCB 104D - KEYPAD / DISPLAY BOARD, 3/3 2 2 uF 0 .1 u F GN D IN T GN D PAD 3 PAD 4 PCF8 5 7 4A PN LED+ LED- U1 A U1 C MM7 4H C1 4 N MM7 4H C1 4 N 2 .7 OH M, 2 W U1 B...
SPARE, 0- 10V PW, 0 -1 0V OFFSET, 0-1 0 V AMPLI TU DE, 0 -10 V TRIG OUT CH S GND Title AV-107C-B, AV-107D-B CHASSIS WIRING Date Rev ision 24-Mar-2005 P: \circui t s\ AV-107\av-107.ddb - AV-107D-B-PN\ POWER and WIRING mod8.sch...
Connecting resistors in parallel reduces the total effective resistance and inductance. Avtech does not recommend the use of single, high-power resistors, even if they are low-inductance types. These resistors experience high failure rates in pulsed applications.
Need help?
Do you have a question about the AV-107C-B and is the answer not in the manual?
Questions and answers