Page 1
Motova 300 English Translation of the original instructions...
Page 2
This symbol is printed on the large mixing bowls. It indicates that these relatively heavy mixing bowls must be held firmly once mixing is complete. BEGO Bremer Goldschlägerei Wilh. Herbst GmbH & Co • Wilhelm Herbst-Straße 1 • D-28359 Bremen Telefon +49 421 2028-0 • Telefax +49 421 2028-100 • www.bego.com...
Motova 300 Safety instructions Motova 300 is designed exclusively for the machine are not permitted. • Always keep signs mixing of investment materials, silicone, and stickers such that they are easy to read. They plaster or other forming materials in dental must not be removed.
Page 5
Motova 300 Special safety instructions concerning this unit z The mixing bowl is held onto the device by vacuum. If the vacuum is insufficient, the mixing bowl will fall down! The mixing bowl must therefore be held firmly after the start of...
Motova 300 Unit description Motova 300 is a vacuum mixing device for the mixing of investment materials, silicone, plaster and other forming materials in the dental laboratory. Various mixing B E GO M o t o v a 3 0 0 programs are available for this purpose.
(see the safety instructions). Switch on Switch on the device with the main switch. The display shows BEGO Motova 300 and the Version-No. of the software. After approx. 3 sec. B E G O M o t o v a 3 0 0...
Motova 300 Filling z Select the appropriate mixing bowl: a) Mixing bowls in the M-series: for investment material and plaster, no silicone Mixing bowls in the MS-series: only for silicone. b) The filling level must lie between the min and max markings! The markings are different in the M and MS mixing bowls.
Motova 300 Mixing Push the mixing bowl with cover into the holder 1 and hold firmly with the hand! P 0 1 M i x Press the START key. B a c k S T A R T A vacuum is then created.
Motova 300 Replace air filter After 999 mixing processes (cycles) the display shows Change filter. The air filter should be replaced but work can be continued immediately: %(*2 0RWRYD Press the STOP key: 1 x mixing Switch the device off and then back on again: 10 x mixing z Remove filter sleeve 1 from the device.
Motova 300 Replace O-rings and discs in mixing bowl Tools such as probes and screwdrivers can be used to replace the O-rings if they are difficult to access. Components must be removed to replace the O-rings and discs S (see diagram).
Motova 300 Elimination of faults Faults Cause Rectification Always switch the device off beforehand! Main switch is No power. Check mains connection. switched on, Remove fuse holder 1 Display is not lit up. Fuse is defective. and replace fuse. If the fuse fails again, contact customer services.