Preparation Before Operation

Environment Demand
The surgical unit should be kept away from other electronic device in the operating room.
The surgical unit may have influence on other electronic device at power output.
Do not put any item on the surgical unit or put surgical unit on any item, so that to ensure
heat dissipation.
Accessories selecting and Installation
Connection of Power Cord: Connect one end of the power cord with the machine, and
then insert the plug to the 3-hole socket
Accessories selecting (according to the mode of power output):
Selecting the hand control pencil output: Insert the plugs of the hand control pencil and
the plate into the desired place directly.
Selecting the foot control pencil output: Insert the plugs of the foot control pencil and the
plate into the desired place directly; insert the plug of pedal switch into the rear panel.
Patient Plate Using
Patient Plate Classification
Two kinds of patient plates are offered, such as hard patient plate and soft patient plate.
Hard patient plate only offers a single patient plate and can be used for many times.
Soft patient plate includes two kinds, such as soft single patient plate and soft dual patient
plate
.
Hard Patient Plate Using
Insert the plug into the desired place, and the LED for single patient plate mode will
illuminate. One must wrap a wet tower soaked in health salt water to the plate, and place it
under the buttocks or leg of patient, so as to keep reliable contact with patient skin. If the
hard patient plate cord is disconnected during the operation, the alarm LED of the machine
will illuminate, the alarm sound will be heard when activation, and the power output will be
interrupted.
Single Soft Patient Plate Using
Insert the plug of patient plate cord into the desired place, and the LED for single
patient plate mode will illuminate. Clip the clamp on the other end of the cord with the
single soft patient plate. Then the patient plate is evenly attached to the patient skin. If the
patient plate cord is disconnected during the operation, the alarm LED of the machine will
illuminate, the alarm sound will be heard when activation, and the power output will be
interrupted.
Heal Force EB03
Operation manual
11