6) Select "DICOM CONFORMANCE TEST" from MENU 1/2 using the "▼" or "▲" button
and press the "OK" button. Then select "Yes" for "START TESTING" from "DICOM
CONFORMANCE TEST" OSD and press the "OK" button.
If "DICOM GSDF" is NOT selected, the message "START TESTING" will not
Note
appear.
START TESTING
Completing the measurement takes between 1 minute to 2 minutes.
7) Showing the test results
When the measurement is completed, the test results are shown.
TEST RESULTS
D I C O M
C O N F O R M A N C E
R E S U L T
M A X
D E V I A T I O N
L U M I N A N C E
E L A P S E D
T I M E
S T A R T
T E S T I N G
Y e s
> N o
The test results are saved until the next test or calibration.
If the achieved maximum deviation does not conform to the standard even though
the test has been performed under the recommended condition, calibrate the monitor
using an external calibration sensor.
Without the OSD operation for over 15 seconds, the OSD disappears automatically.
DICOM CONFORMANCE TEST
D I C O M
C O N F O R M A N C E
R E S U L T
M A X
D E V I A T I O N
L U M I N A N C E
E L A P S E D
S T A R T
T E S T I N G
> N o
T E S T
:
P a s s e d
:
4 . 5
%
:
S t a b l e
:
1 2 3 m i n
- 15 -
T E S T
:
- - -
:
- - -
:
S t a b l e
T I M E
:
1 2 3 m i n
Y e s
Test results
Passed: Max.
deviation ≤ 10 %
Failed: Max. deviation > 10 %
Cancelled: Test has been
5 M P
cancelled.
Passed(!), Failed(!)
When "Passed" or "Failed" is
followed by "(!)", the test has
been made under not
recommended condition.
(The test was performed in less
than 60 minutes of continuous
lighting time of the backlight.)
If the test is continuously
performed in less than 30
seconds, "Unstable" is shown in
LUMINANCE and "(!)" is shown in
RESULT.
Maximum deviation
5 M P