Download Print this page

HP 416A Instructions And Operating Manual page 66

Ratio meter

Advertisement

Tube
Vl8
Vl9
czo
V21
Function
Rectifier, Power Supply Circuit.
Series Regulator Tube, Voltage
Regulator Circuit.
Control Tube, Voltage Regu
-
lator Circuit.
Voltage Reference Tube
,
Volt
-
age Regulator Circuit.
Required Circuit Adj. (cont.)
Adjustment of HV Output
Level and Check of Line
Regulation.
Adjustment of HV Output
Level and Check of Line
Regulation.
Adjustment of HV Output
Level and Check of Line
Regulation.
Adjustment of HV Output
Level and Check of Line
Regulation.
Capacitors
-
The electrolytic capacitors
in
The Model 416A are high
quality capacitors which have a useful life of from five to ten years.
Do not
replace these capacitors unless they are proven defective by accurate tests.
Variable Resistors
-
Should any of the variable resistors require re
-
placement,
connections should be made as indicated in Fig
Q
19
0
The adjust
-
ment to be performed if a
variable
resistor
is
replaced is described in the
Adjustments section.
If R60 is replaced, readjust the tuning eye (V9)
.
If R94 or R96 is replaced
,
recalibrate the meter for electrical zero
and full
-
scale deflections
Q
If Rll6 is replaced
,
readjust the regulator output
voltage
leveL
If Rl 18 is replaced
,
readjust the
filament
supply voltage leveL
Selenium Rectifier
-
Should rectifier SR 1 need replacement
,
connec
-
tions should be made as indicated in Fig. 26
.
-