Releasing the locking brakes

WARNING
Risk of personal injury and/or damage to the
device
Fully equipped (= heavily laden) ceiling supply
units require high braking forces due to the
inertia of the weight. The braking force must
be applied by the user because the ceiling
supply unit may not be stopped with the lock-
ing brake.
The bearings on the arm systems are always fitted
with friction brakes. Additionally, pneumatic brakes
may also be present.
Friction brakes
 Arm systems with friction brakes can be posi-
tioned without further preparation, see "Posi-
tioning the ceiling supply unit" on page 22.
Pneumatic brakes
CAUTION
Damage to the pneumatic brake
Systems with a pneumatic brake must not be
turned with the locking brake engaged. The brake
can be damaged as a result.
The system must not be moved without the brake
being released and the key may only be released
after the system has come to a standstill. Move-
ments of the support arm may only be braked by
hand.
 For arm systems with a pneumatic brake, the
brake must be released using the keys.
The keys are installed in the media head or media
column.
Instructions for use Agila CC/Agila CC lift express
NOTE
To release the pneumatic brake on a bearing, it is
necessary to press the key that is labeled with the
corresponding bearing.
At the ceiling bearing
1 Hold down key (A) and move the arm system
and media column/media head (optional) at the
same time.
2 Bring the arm system and media column/media
head (optional) into the desired position, see
"Positioning the ceiling supply unit" on page 22.
Only release the key (A) once the system has
come to a standstill. The arm system is locked
again.
At the intermediate bearing
1 Hold down key (B) and move the arm system
and media column/media head (optional) at the
same time.
2 Bring the arm system and media column/media
head (optional) into the desired position, see
"Positioning the ceiling supply unit" on page 22.
Only release the key (B) once the system has
come to a standstill. The arm system is locked
again.