Series TC-Standard Contents Introduction ............................................. 3 Intended use ......................................... 3 Overview .......................................... 3 Scope of delivery ........................................ 3 Ordering instructions...................................... 4 1.4.1 Gas cooler models with one heat exchanger .......................... 4 1.4.2 Gas cooler models with one heat exchanger for H2/O2 applications .................. 5 1.4.3 Gas cooler models with two heat exchangers .......................... 6 1.4.4...
Page 4
Series TC-Standard 7.10.2 Spare parts and accessories for cooler with -H2/-O2 heat exchanger................ 39 8 Disposal ............................................ 40 9 Appendices ............................................. 41 Gas cooler technical data .................................... 41 Technical Data - Options .................................... 44 Flow diagrams ........................................ 45 Performance curves ...................................... 46 Heat exchanger .........................................
1.2 Overview The TC-Standard series was designed specifically for high cooling capacities and high ambient temperatures. The TC-Standard+ series was designed specifically for the requirements of automated measuring systems (AMS) according to EN 15267-3. The series connection of the heat exchangers will cool in two cycles to minimise wash-out effects.
Series TC-Standard 1.4 Ordering instructions 1.4.1 Gas cooler models with one heat exchanger The item number is a code for the configuration of your unit. Please use the following model code: 4496 2 1 1 X 0 X 1 X X X 0 X X X 0 X 0 0 Product Characteristics Gas cooler models (with 1 heat exchanger) TC-Standard 6111: Ambient temperature 40 °C TC-Standard 6112: Ambient temperature 50 °C...
Series TC-Standard 1.4.2 Gas cooler models with one heat exchanger for H2/O2 applications The item number is a code for the configuration of your unit. Please use the following model code: 4496 2 1 1 X 0 X 1 1 X 0 0 0 0 X 0 X 0 X Product Characteristics Gas cooler models (with 1 heat exchanger) TC-Standard 6111: Ambient temperature 40 °C TC-Standard 6112: Ambient temperature 50 °C...
Series TC-Standard 1.4.3 Gas cooler models with two heat exchangers The item number is a code for the configuration of your unit. Please use the following model code: 4496 2 1 2 X 0 X 2 X X X 0 X X X 0 X 0 0 Product Characteristics Gas cooler models (with 2 heat exchangers) TC-Standard 6121: Ambient temperature 40 °C TC-Standard 6122: Ambient temperature 50 °C...
Series TC-Standard 1.4.4 Gas cooler model with two heat exchangers in series The item number is a code for the configuration of your unit. Please use the following model code: 4496 2 1 2 X 0 X 1 X X X 0 X X X 0 0 0 0 Product Characteristics Gas cooler models (with 2 in-line heat exchangers) TC-Standard+ 6121: Ambient temperature 40 °C TC-Standard+ 6122: Ambient temperature 50 °C...
Series TC-Standard 2 Safety instructions 2.1 Important advice Operation of the device is only permitted if: – the product is used under the conditions described in the installation- and operation instruction, the intended application according to the type plate and the intended use. In case of unauthorized modifications done by the user Bühler Technolo- gies GmbH can not be held responsible for any damage, –...
Series TC-Standard 2.2 General hazard warnings The equipment must be installed by a professional familiar with the safety requirements and risks. Be sure to observe the safety regulations and generally applicable rules of technology relevant for the installation site. Prevent malfunctions and avoid personal injuries and property damage.
Series TC-Standard 3 Transport and storage Only transport the product inside the original packaging or a suitable alternative. The equipment must be protected from moisture and heat when not in use. It must be stored in a covered, dry and dust-free room at a temperature of -20 °C to 60 °C (-4 °F to 140 °F).
Series TC-Standard 4 Installation and connection 4.1 Installation site requirements The unit is only intended for wall-mounted use in enclosed areas. Adequate protection from the weather must be provided when used outdoors. Install the unit leaving enough room below the cooler to discharge the condensate. Leave room above for the gas supply. Be sure to maintain the approved ambient temperature.
Series TC-Standard 4.2.3 Connecting the moisture detector (option) When ordering the cooler with moisture detector option, it will be factory installed inside a flow adapter, or for the filter option installed and connected in the filter head. 4.2.4 Peristaltic pump connector (optional) Coolers ordered with attached peristaltic pump already have it installed and wired.
Series TC-Standard On glass heat exchangers, the correct position of the seal is important when connecting the gas lines (see image). The seal con- sists of a silicone ring with a PTFE sleeve. The PTFE side must face the glass thread. Glass PTFE Pay attention to the appropriate spanner size when selecting fittings for stainless steel heat exchangers.
Page 16
Series TC-Standard WARNING High voltage Damage to the device in case of insulation testing Do not proceed insulation tests with high voltage to the device as a whole! Electric strength test This device is equipped with extensive EMC protection. The necessary tests were carried out at the factory (test voltage 2.1 kV or 2.55 kV DC depending on approval).
Series TC-Standard 4.4 Signal outputs The device provides different status signals. The maximum switching load of the alarm outputs is limited to 250 V AC/150 V DC, 2 A, 50 VA each. An alarm is triggered by the alarm contact/status output (S2) if the temperature of the cooler is outside the specified limits. It does not indicate if the alarm was triggered due to excess temperature or insufficient temperature.
Series TC-Standard 5 Operation and control NOTICE The device must not be operated beyond its specifications. After switching on the cooler the block temperature will be displayed. The display will flash until the block temperature has reached the preset target value (± adjustable alarm range). The status contact is in the Alarm position. Once the target temperature range has been reached, the temperature will continuously be displayed and the status contact switches over.
Series TC-Standard 5.3 Use of menu functions Brief description of the operating principle: The unit is operated using 5 keys. Their functions are: Button Section Functions Display – Switches from the measurement display to the main menu Menu – Selects the menu item displayed Enter –...
Series TC-Standard 5.3.2 Menu navigation overview When pressing the OK button in normal mode, the display will show the prompt code if the menu is locked. Use the ▲ and ▼ buttons to enter the correct code and press OK. If an incorrect code or no code is entered, the menu will not be unlocked and you will not be able to access all menu items.
Page 21
Series TC-Standard Display Main menu Submenu 1 Submenu 2 Parameter tcSt tEmP 5°C Display TC Standard Temperature Input Temperature CH1 Device designation Target temp. Cooling block 2 . . . 20 °C A hi 3°C Display Settings menu Input Alarm hysteresis 1 .
Series TC-Standard 5.4 Description of menu functions 5.4.1 Display menu Block temperature display Display → Depending on the device state, the temperature will be displayed as a constant, flashing, or alternat- ing with a status message. Ambient temperature display Display → The display is only available on devices with „Delta-T"...
Series TC-Standard Peristaltic Pump Display → PuMP Switching the peristaltic pump on and off. Parameter range: Factory setting: Note: Status switches, "PuMP" flashes. Heat exchanger material selection Display → h. c hn Heat exchanger material selection Parameter range: (Steel), (Glass), (PVDF) Factory setting: (cooler without heat exchanger), or respective material per configuration...
Series TC-Standard lower alarm limit Display → Cooler → A Lo (Alarm low) Here you can set the lower threshold for the visual signal and the alarm relay. The alarm limit is set based on the cooler temperature setting. Parameter range: -1 °C to -3 °C (-1.8 °F to -5.4 °F) Factory setting: -3 °C (-5.4 °F)
Page 25
Series TC-Standard Calibrate moisture detector Display → → If a moisture detector is installed, calibration can now be performed. To do so, the unit must be flushed with dry gas. Note: Calibration was performed at the factory using ambient air. After replacing the moisture detector a calibration is again required.
Page 26
Series TC-Standard Lock Menu To protect the menu from unauthorised use, enter a value for the lock code. Menu items can then only be accessed after enter- ing the correct code. Display → → This setting will cancel/enable the menu lock. Parameter range: 0 to 9999 Factory setting:...
Page 27
Series TC-Standard Modbus RTU device ID Selecting the device ID for communication via digital interface. The ID can be any within a defined range, the default is 10. The properties of the digital interface are not affected when resetting the device to its factory settings! This menu item is only available on devices with "Digital output Modbus RTU"...
Series TC-Standard 5.4.4.1 Submenu 2 (Analog Output 1) The analog output will display the actual cooler temperature. The menu items for the analog output are no available on devices with "Digital output Modbus RTU" option. Signal behaviour In normal mode (noP) the measuring point will output the actual temperature. For testing purposes you can generate constant values hi, or hALF.
Series TC-Standard 5.5 Using the Digital Interface The digital interface on this device is a Modbus RTU protocol, which physically communicates via RS485 (2-wire). The cooler therefore takes on the role of the slave in communication. The Modbus interface enables direct access to process and diagnostic data and parameters during operation. 5.6 Modbus Configuration The settings below are the defaults;...
5.8 Modbus Register Description Address Access Data type Default Selection Resolution Unit Block temperature measurement 2000 Float °C Block temperature status 2002 Uint32 Bit 0 := Error Bit 1..15 := Reserved Bit 16:= Sensor not calibrated Bit 17:= Initialization/measurement invalid Bit 18 := Stabilisation phase Bit 19:= Load limit reached Bit 20:= Measurement outside target range...
Page 31
Description Address Access Data type Default Selection Resolution Unit Heat exchanger type selection 3, 16 9007 Uint16 2 := Steel 3 := Glass 4 := PVDF De-/activate condensate pump(s) 3, 16 9008 Uint16 3:= Pump active 4:= Pump inactive Modbus Baudrate selection 3, 16 9009 Uint16...
Page 32
Description Address Access Data type Default Selection Resolution Unit Bit 5 := Bit 6 := FF1 moisture alarm Bit 7 := FF2 moisture alarm Condition code register 3 10003 Uint16 Bit 0 := Peristaltic pumps deactivated Bit 1 := Bit 2 := Bit 3 := Bit 4 := Bit 5 :=...
Page 33
Description Address Access Data type Default Selection Resolution Unit Bit 3 := Measurement fluctuation Bit 4 := Bit 5 := Bit 6 := Bit 7 := Error register 6- PT100.2 10010 Uint16 Error register 7 10011 Uint16 Error register 8 10012 Uint16 Error register 9...
Series TC-Standard 6 Maintenance The base version of the cooler requires no special maintenance. However, it may have different options depending on the cooler model. In this case the following routine maintenance is re- quired: – Option peristaltic pump: Check hoses –...
Series TC-Standard 7 Service and repair This chapter contains information on troubleshooting and correction should an error occur during operation. Repairs to the unit must be performed by Bühler authorised personnel. Please contact our Service Department with any questions: Tel.: +49 (0) 2102-498955 or your agent For more information on your custom maintenance and startup services, please visit: https://www.buehler-technologies.com/ service If the device is not functioning properly after correcting any malfunctions and switching on the power, it must be inspected by...
Series TC-Standard 7.1.1 Error messages on the display If an error occurs, the display will read "Err". Press the " " button to show the error number(s). Error messages will appear until the unit has been restarted or the error is cleared using the "Func" button. It can only be cleared if the cause for the error has been corrected.
Series TC-Standard CAUTION Health hazard if the heat exchanger leaks The heat exchanger is charged with glycol-based coolant. In the event of a heat exchanger leak: a) Avoid contact with the skin and eyes. b) In the event of a leak, do not restart the cooler under any circumstances The cooler must be repaired by the manufacturer.
Series TC-Standard 7.5 Replacing the hoses of the peristaltic pump (option) – Turn off gas supply. – Switch the device off and disconnect power supply. – Remove the supplying and draining hoses from the pump (Take care of the safety instructions!). –...
Series TC-Standard 7.8 Calibration of the moisture detector (option) – When replacing the moisture detectors, they must be recalibrated. – Be sure dry gas flows through the cooler. – Select cooler menu and confirm. – Select menu item moisture detector. –...
Series TC-Standard 8 Disposal The heat exchanger is charged with glycol-based coolant. The applicable national laws must be observed when disposing of the products. Disposal must not result in a danger to health and environment. The crossed out wheelie bin symbol on Bühler Technologies GmbH electrical and electronic products indicates special disposal notices within the European Union (EU).
Series TC-Standard 9 Appendices 9.1 Gas cooler technical data TC-Standard Gas Cooler Technical Data Ready for operation after max. 10 minutes Ambient temperature 5 °C to 50 °C Gas outlet dew point preset: 5 °C adjustable: 2 °C…20 °C or Delta T control IP rating IP 20 Mechanical load...
Page 44
Series TC-Standard TC-Standard with -H2/-O2 heat exchanger Gas cooler technical data Ready for operation after max. 10 minutes ambient temperature 5 °C to 50 °C Gas outlet dew point preset: 5 °C adjustable: 2 °C…20 °C or Delta T control IP rating IP 20 Mechanical load Tested based on DNV-GL CG0339 vibration class A (0.7 g) 2 Hz-13.2 Hz amplitude ±...
Page 45
Series TC-Standard TC-Standard+ Gas Cooler Technical Data Ready for operation after max. 10 minutes Ambient temperature 5 °C to 50 °C Gas outlet dew point preset: 5 °C adjustable: 2 °C…20 °C IP rating IP 20 Mechanical load Tested based on DNV-GL CG0339 vibration class A (0.7g) 2 Hz-13.2 Hz amplitude ±...
Series TC-Standard 9.2 Technical Data - Options Analogue Output Cooler Temperature Technical Data Signal 4-20 mA or 2-10 V corresponds to -20 °C to +60 °C cooler temperature Connection M12x1 plug, DIN EN 61076-2-101 Digital interface technical data Signal Modbus RTU (RS-485) Connection M12x1 connector, DIN EN 61076-2-101 Technical Data FF-3-N Moisture Detector Ambient temperature 3 °C to 50 °C...
Series TC-Standard 9.4 Performance curves TC-Standard One heat exchanger Two heat exchangers Model TC-Standard 6111 Model TC-Standard 6121 Rated cooling capacity (at 25 °C) 100 kJ/h Rated cooling capacity (at 25 °C) 100 kJ/h Max. Ambient temperature 40 °C Max. Ambient temperature 40 °C Dew point fluctuations Dew point fluctuations...
Page 49
Series TC-Standard TC-Standard with -H2/-O2 heat exchanger Model TC-Standard 6111 Model TC-Standard 6112 Rated cooling capacity (at 25 °C) 100 kJ/h Rated cooling capacity (at 25 °C) 90 kJ/h Max. ambient temperature 40 °C Max. ambient temperature 50 °C Dew point fluctuations Dew point fluctuations static ±...
Series TC-Standard 9.5 Heat exchanger 9.5.1 Heat exchanger description The energy content of the sample gas and the required cooling capacity of the gas cooler is determined by three parameters: gas temperature ϑ , dew point τ (moisture content) and volume flow v. The outlet dew point rises with increasing energy content of the gas.
Page 51
Series TC-Standard TC-Standard+ Heat exchanger 2x MTG-2 2x MTV-2 2) 3) 2) 3) 2x MTG-2-I 2x MTV-2-I Materials in contact with media Glass PVDF PTFE Flow rate v 210 Nl/h 190 Nl/h Inlet dew point τ 70 °C 70 °C e,max Gas inlet temperature ϑ...
Series TC-Standard 10 Attached documents – Declaration of Conformity KX440006 – O declaration (heat exchanger) – RMA – Decontamination Statement BE440017 ◦ 03/2024 Bühler Technologies GmbH...
Page 58
Heat exchanger optimized for use with high-purity oxygen Gas Analysis Applications with oxygen: Free from particles, oil and grease For use with high-purity oxygen, the product requires special cleaning to ensure that it is free from oil and grease, as oxygen is a strong oxidising agent. Under unfavourable conditions, oxy- gen can cause spontaneous combustion of organic substances such as particles, oils and fats, and generally promotes the combustion of substances.
Page 59
RMA-Formular und Erklärung über Dekontaminierung RMA-Form and explanation for decontamination RMA-Nr./ RMA-No. Die RMA-Nr. bekommen Sie von Ihrem Ansprechpartner im Vertrieb oder Service. Bei Rücksendung eines Altgeräts zur Entsorgung tragen Sie bitte in das Feld der RMA-Nr. "WEEE" ein./ You may obtain the RMA number from your sales or ser- vice representative.