Download Print this page

Technical Specifications; Emc Information - homed JLN-S0 A Series User Manual

Piston compressor nebulizer

Advertisement

Shenzhen Homed Medical Device Co.,Ltd
- Take off the cover of the compartment which houses the air filter;
- Remove the old filter and insert a new one;
- Replace the cover.
- Do not use cotton or any other materials. Do not wash or clean the filter. Use only filters provided by
HOMED and don't operate without a filter.

8 Technical specifications

- The maximum A-weighted sound pressure level: less than 55dBA
- Motor: Compact piston compressor BD80 CX
- Power supply: AC 100-240V 50/60Hz
- Working pressure: W
- Air flow: 8.0 l/min
- Air filter included
- Maximum nebulizing rate: ≥0.15ml/min
- Operating Temperature Range: 50℉ to 104℉ (10℃ to 40℃)
- Operating Humidity Range: 10 to 95% RH
- Storage Temperature Range: -40℃ to 70℃
- Storage Humidity Range: 10 to 95 % RH
-Transportation conditions: -40℃-70℃, 10%-95%RH
- Dimension: 95 x 135 x 55 mm
- Weight: 425 g
- Waterproof degree: IPX0
- Degree of safety in the presence of flammable anesthetics or oxygen: No AP/APG (Not suitable for use in the
presence of flammable anesthetice or oxygen)

9 EMC information

This medical device manufactured by HOMED conforms to this EN60601-1-2:2007 standard for both
immunity and emissions.
Nevertheless, special precautions need to be observed:
Note: Do not use mobile (cellular) telephones and other devices which generate strong electrical or
electromagnetic fields, near the medical device.
This may result in incorrect operation of the unit and create a potentially unsafe situation. Recommendation
is to keep a minimum distance of 7 m. Verify correct operation of the device in case the distance is shorter. See
the table 9.1 to table 9.4 in the following:
JLN-S0xA series piston compressor nebulizers are intended for use in the electromagnetic environment
specified below; The customer or the user should assure that it is used in such and environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
ithout tube and atomizer: 70 Kpa -100 Kpa/0.7-1bar;
Table 9.1 Electromagnetic emissions
Compliance
Group 1
Class B
Not
applicable
Electromagnetic environment-guidance
JLN-S0xA uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
File No.: JLN-S0XA-CE-14
w
ith tube and atomizer: at least 0.19Mpa
Page 6 of 10

Advertisement

loading

This manual is also suitable for:

Jln-s01aJln-s02aJln-s03aJln-s04 aJln-s05aJln-s06a