INTRODUCTION
This instrument is a 4 1/2 digit, compact-sized portable digital cali-
brator thermometer designed to use external K/J/T/E/R/S/N/L/U/
B/C type thermocouples as temperature sensor. The thermometer
features a dual thermocouple input, an adjustable T/C offset. The
thermocouples types comply with the(N.I.S.T. Monograph 175 Re-
vised to ITS 90 standard).
WARNING
To avoid electrical shock, do not use this instrument when work-
ing voltages at the measurement surface over 24V AC or DC.
WARNING
To avoid damage or burns, do not make temperature measurement
in microwave ovens.
CAUTION
Repeated sharp flexing can break the thermocouple leads. To pro-
long lead life, avoid sharp bends in the leads, especially near the
connector.
SAFETY INFORMATION
It is recommended that you read the safety and operation instruc-
tions before using the thermometer.
GENERAL SPECIFICATIONS
Displays:
There are three displays Main Second and Third. The Main and
Second display panels are 4 1/2 digit liquid crystal display (LCD)
with maximum reading of 19999, The main is used for displaying
the value of T1, T2 or output setting. The second displays T1 or T2
readings and the third T1-T2 and groups settings.
Battery:
Standard 9V battery (NEDA 1604, IEC 6F22 006P). Battery life is
about 17.5 hours when used with a carbon zinc battery.
Low battery indication:
The "
" is displayed when the battery voltage drops below the
operating level.
Dimensions:
192mm(H) x 91mm(W) x 52.5mm(D).
Weight:
318g.
Accessories:
Two type "K" thermocouple bead wires. Two type "K" thermocou-
ple calibration bead wires. Maximum insulation temperature 260°C
(500°F). Wire accuracy ±2.2°C or ±0.75% of reading (whichever is
greater) from 0°C to 800°C.
A 9 volts battery.
An instruction manual.
ENVIRONMENTAL
Ambient Operating Ranges:
0°C to 50°C (32°F to 122°F) <80% R.H.
Storage Temperature:
-20°C to 60°C (-4°F to 140°F) <70% R.H.
Input Connector:
Accepts standard miniature thermocouple connectors (flat blades
spaced 7.9mm, center to center).
SPECIFICATIONS
ELECTRICAL
Temperature Scale:
Celsius or Fahrenheit user-selectable.
Measurement
Range:
K-TYPE(0.1°) -200°C to 1372°C or -328°F to 2501°F
J-TYPE(0.1°) -210°C to 1200°C or -346°F to 2192°F
T-TYPE(0.1°) -250°C to 400°C or -418°F to 752°F
E-TYPE(0.1°) -250°C to 1000°C or -418°F to1832°F
R-TYPE(1°) 0°C to 1767°C or 32°F to 3212°F
S-TYPE(1°) 0°C to 1767°C or 32°F to 3212°F
N-TYPE(0.1°) -200°C to 1300°C or -328°F to 2372°F
L-TYPE(0.1°) -200°C to 900°C or -328°F to 1652°F
U-TYPE(0.1°) -200°C to 600°C or -328°F to 1112°F
B-TYPE(1°) 600°C to 1820°C or 1112°F to 3308°F
C-TYPE(1°) 0°C to 2316°C or 32°F to 4200°F
Based on the ITS-90 temperature standard.
According to temperature standard ITS-90.
Accuracy:
K/J/T/E/L/U-TYPE
±(0.05% rdg + 0.5°C) -50°C to 1372°C
±(0.05% rdg + 1.0°C) -50°C to -250°C
±(0.05% rdg + 1.0°F) -58°F to 2501°F
±(0.05% rdg + 2.0°F) -58°F to -346°F

Advertisement

loading

Summary of Contents for TECPEL CL 326A

  • Page 1 INTRODUCTION Accessories: Two type “K“ thermocouple bead wires. Two type “K” thermocou- CL 326A This instrument is a 4 1/2 digit, compact-sized portable digital cali- ple calibration bead wires. Maximum insulation temperature 260°C brator thermometer designed to use external K/J/T/E/R/S/N/L/U/ DIGITAL THERMOMETER (500°F).
  • Page 2 N-TYPE Press ““ button to make left shifts to cycle through K→mV→C new minimum or maximum measurement is recorded. Press the MIN/MAX button again to rotate through the current →B→U→L→N→S→R→E→ T→J. ±(0.05% rdg + 1.0°C) -50°C to 0°C readings: Press the “type” button again to choose the selected thermocou- ±(0.05% rdg + 0.5°C) 0°C to 1300°C MAX: The highest measurement recorded.