Hide thumbs Also See for 870:

Advertisement

Quick Links

Open\Display\Ghartgraph
Button
Description
[Exit]
To exit TECPEL 870 application.
Ⅲ Setup, Start & Exit on Software
Quick Start at the TECPEL 870
1. Install the RS-232 cable between PC's Com port and TECPEL 870's RS-232 port.
Please make sure:
- The RS232 cable connector (to meter) faced up with panel.
- Physical Com port number has to be the same as setup in software.
2. Turn the TECPEL 870 on.
At the PC
1. Open the TECPEL 870 software by double-clicking on the icon SETUP.exe.
2. Please double-clicking TECPEL 870 on the file C:\TECPEL 870\ to start the TECPEL 870
application.
C. Quick Exit
1. Click Exit to exit TECPEL 870 application.
10
Operation Manual
TECPEL
870

Advertisement

loading

Summary of Contents for TECPEL 870

  • Page 1 2. Turn the TECPEL 870 on. At the PC 1. Open the TECPEL 870 software by double-clicking on the icon SETUP.exe. 2. Please double-clicking TECPEL 870 on the file C:\TECPEL 870\ to start the TECPEL 870 application. C. Quick Exit...
  • Page 2: Table Of Contents

    2. Change the path if necessary or choose “Finish” button to install program directly. 2. General ..........................1 Note: 3. Specifications .........................1 1. Check the TECPEL 870 software disk for virus before installation. 4. Front Panel Description ......................2 2. System required: 80486-33 with 16MB RAM or better. 4.1 Push Button functions......................2 Ⅱ...
  • Page 3: Introduction

    9. Trouble Shooting 1. Introduction ? Power on but no display: Check the battery are in place and making good contact or correct This instrument is a portable easy to use 3 ½ digit, compact-sized pH/mV and temperature me- polarity, replace a new battery. ter.
  • Page 4: Front Panel Description

    4. Front Panel Description 7. Measuring Procedure 7.1 pH Measurement Calibrate the instruments and pH electrode before measuring. 1. Connect the combination pH electrode to the BNC socket. 2. Power on the instrument by pressing the “MEAS” button. 3. If the operation is under the “ATC” then please refer 5.1.1 Automatic Temperature com- pensation mode.
  • Page 5: Meas" Button

    6.3 Temperature Calibration: 4.1.4 “MEAS” Button: 1. Select the function in mV and connect the temperature plug. It can turn on or off the meter by pressing “MEAS” button, power is automatically turn 2. Sliding the “CAL” switch to right to enter calibration procedure. off if no operation for a period of 10 min.
  • Page 6: Ph Calibration With Manual Temperature Compensation(Mtc)

    5. Press and hold down “SET” button for more than 2 seconds to enter the pH4.01 mode, 5. Press and hold down “SET” button for more than 2 seconds to enter the pH4.01 calibra- and the main (pH) display blanks. (Fig-3) tion procedure, and the sub(Temp.) display blanking.