Installation manual of AA6688 Operating Table P/N:8501191 version:C date of issue:2021-2-25...
Page 3
Copyrights ADS makes no warranty of any kind with regard to this material , including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. ADS shall not be held liable for any errors contained herein or any consequential or other damages concerning furnishing, performance or use of this material.
Installation manual of Operating Table 1 Introduction Attention!This Instruction is ensure for the users use the Operating Table correctly. Please read the manual carefully before installation and operation; Strictly observe the precautions in the instruction manual, otherwise it may cause injury to patients or operators.
Installation manual of Operating Table 1.5 Specifications List Power AC110-240V 50/60Hz Output power 420VA Load capacity 135kg Highest chair position 1020mm Lowest chair position 620mm Backrest adjustment angle range 0~85° Feet adjustment angle range 0~80° Headrest telescoping length 150mm Head oblique angle -30°...
Installation manual of Operating Table Warning Please read this manual carefully before installing and using an electric operating table. Ø It is recommended that the product installation process be performed by two people. Ø It is recommended to place the product on a clean, level floor or platform. Ø...
Installation manual of Operating Table 4 Installation steps Force-bearing point 1. Remove the packing box Unpack all packages and protective items of the operating table; Loosen 8 screws of the base fixed bracket and remove the bracket; 2. Placing the operating table Remove the packing of the operating table, grasp the force point, lift the operating table and place it on the flat ground. 3. Insert the rod of headrest to the hole...
4、Put the left and right supporting hands upright respectively, to make the fixed column is inserted into the mounting hole of the backrest, and then rotate and put down after it is inserted in place. 5、 Hang the Manual Operation Panel to the pole near armrest,put the foot pedal to the suitable place; connect the power and test various functions of Operating Table. foot controller 6、Step on the green end of the pedal at both ends, and the caster is in a movable state. Push the chair to a suitable position in the treatment room, and then step down the red end of the foot pedal at both ends to make the caster in a locked state and the operating table is fixed. Notice In order to fix the operating table completely, the casters at both ends must be locked;...
Installation manual of Operating Table 5 Control function(Software uersion:1.0.0) Foot pedal introduction Operating Table down Cushion down Operating Table up Cushion up Backrest down Stop Backrest up Program Feet support cushion down Reset Programmable positions Feet support cushion uo (P1/P2/P3/P4)...
Installation manual of Operating Table 5.2 Foot pedal function introduction 、Operating Table down:control Operating Table down When press the button(not let go), M1 motor that controls the Operating Table up and down contracting ,Operating Table down;when loose the button, M1 motor will stop。 2、...
Page 15
Installation manual of Operating Table 5、Foot support cushion down:control foot support cushion down When press the button(not let go), M3 motor that controls the foot support cushion contracting, foot support cushion down; when loose the button,M3 motor will stop. 6、 Foot support cushion up:control foot support cushion up When press the button(not let go), M3 motor that controls the foot support cushion reaching out,...
Page 16
Installation manual of Operating Table 9、Stop: Press the button,all the program is stop Stop:press this button, buzzer beeps once,all the button function of panel and pedal is useless. Recover:press this button again, buzzer beeps twice, all the button function of panel and pedal is normal;...
Installation manual of Operating Table 5.3 Manual Operation Panel Backrest up Feet support cushion down Programmable positions Backrest down (P1/P2/P3/P4) Operating Table up Program Operating Table down Stop Cushion up Reset Cushion down Power off Feet support cushion up...
Installation manual of Operating Table 5.4 Manual Operation Panel Function 1. backrest up:control backrest up When press the button(not let go), M2 motor that controls the backrest reaching out, backrest up;when loose the button,M2 motor will stop. 2. backrest down:control backrest down When press the button(not let go),green light flashes,M2 motor that controls the backrest contracting,...
Page 19
Installation manual of Operating Table 7. Foot support cushion up:control foot support cushion up When press the button(not let go), green light flashes,M3 motor that controls the foot support cushion reaching out, foot support cushion up; when loose the button,green light on M3 motor will stop 8.
Page 20
Installation manual of Operating Table 11、 Press the button,all the program is stop stop:press this button, buzzer beeps once,all the button function of panel and pedal is useless. recover:press this button again, buzzer beeps twice, all the button function of panel and pedal is normal.
Installation manual of Operating Table 6 Cleaning and Disinfection Warning:please make sure the power is off before cleaning Cleaning: The purpose of cleaning is to remove the stains and dust on the surface of the operating table primarily. It is recommended to clean the surface of operating table with the wet soft cloth and a mild detergent (or the hospital's recommended cleaning solution) once a day after use . with extra attention to areas that may hide dirt or dust. Disinfection: The purpose of disinfection is to remove potentially infectious materials primarily, such as body fluids or used operating tables in patients with known infectious diseases. It is recommended that that 3% hydrogen peroxide should be used after each use to wipe and disinfect the metal parts and plastic parts of the operating table, as well as the leather such as headrest, cushion and armrest. Precautions for disinfection It is strictly forbidden to use alcohol, fortified glutaraldehyde disinfectant, 84 disinfectant, benzalkonium bromide-containing disinfectant, benzsolol ammonium propanol-containing disinfectant and various acid and alkaline disinfectants for disinfection. Chromatic aberration, cracking and aging are not covered by the normal operating table warranty. 7 Circuit diagram Operating table up-down motor Backrest up-down motor Foot support cushion up down motor Cushion up-down motor...
Installation manual of Operating Table 8 EMC attention Attention: • The Operating Table complies with the relevant electromagnetic compatibility requirements of IEC60601-1-2 and IEC60601-2-46 standards. • Users should install and use according to the electromagnetic compatibility information provided in the accompanying documents. •...
Page 23
Installation manual of Operating Table Statement of Manufacturer— Electromagnetic Immunity Operating Table is designed for electromagnetic environment described below. Be sure to apply. Electromagnetic Electromagnetic IEC6061 Test Level Test Level Conformance Immunity Test Environment The floor should be of wood, concrete, or tile. If Electrostatic ±6 kV Contact Discharge±8 kV ±6 contact Discharge±8...
Page 24
Installation manual of Operating Table Statement of Manufacturer— Electromagnetic Immunity Operating Table is designed for electromagnetic environment described below. Be sure to apply. Electromagnetic IEC6061 Test Level Test Level Conformance Electromagnetic Environment Immunity Test The isolation distance between portable and mobile RF 3V(Effective Value) communications devices and any 3V/m part of the Extraoral Dentistry...
Installation manual of Operating Table Recommended isolation distance between portable and mobile RF communication devices and Operating Table Operating Table are intended for use in electromagnetic environments where RF radiation disturbances are controlled. Depending on the maximum rated output power of the communication device, the purchaser or user can prevent electromagnetic interference by maintaining the minimum isolation distance between portable and mobile RF communication devices (transmitters) and Operating Table as recommended below.
Installation manual of Operating Table 11 Explanation of graphics, symbols, abbreviations Store in Keep dry Indoor area Notice and Type B warnings application Upward No rolling Refer to the Serial number manual Grounding Take with care protection Date of Manufacturer manufacture Storage limited Storage limited humidity pressure Storage limited temperature...