ASSEMBLY INSTRUCTION

Equipment commissioning: Make sure the CV-Log-35 is not powered when connecting/disconnecting sensors. A
connection or disconnection of a CV-Log-35 analog input while powered on could damage the product.
1.
Sensor connection
If the CV-Log-35 sensors are not factory wired, start by connecting the desired sensors to the inputs of the product.
Refer to the CVLog3500 wiring diagram for connection details.
2.
Battery connection
Before the CV-Log-35 is commissioned, plug in the battery:
a. Internal battery (standard): open the product by turning the CV-Log-35 head (blue part of the product) anticlockwise
by 1/8 of a turn and connect the battery. Then properly close the product making sure to turn the blue lid clockwise
until stop.
b. High-capacity external battery (option): connect the circular connector of the external battery to the corresponding
connector located on the black part of the CV-Log-35 enclosure.
3.
Enable CV-Log-35
To activate the product, keep the front button
on the front panel of the D35. Release the button after the flash.
2

OPERATING MODE

The CV-Log-35 has 3 modes of operation:
"Standby" mode
"Acquisition" mode
"Configuration" mode
Press the front panel button
In "Standby" mode you can remove the SIM card or SD card, as well as connect or disconnect physical inputs.
The "Acquisition" mode is the operating mode of the CV-Log-35. In this mode, the device acquires the signals from the
connected sensors and saves them to the internal memory. If the data communication option is enabled, the recorded data is
sent across the cellular network at the set interval time.
The "Configuration" mode is used to activate the WiFi local network generated by the device, to configure the CV-Log-35.
Figure 1: Logic diagram showing the devices various modes of operation and how to switch from one to another
 CLA-VAL Europe
© Copyright CLA-VAL Europe - Specifications subject to change without notice - no contractual illustrations.
pressed for approximately 5 seconds until a green LED lights up continuously
to navigate between modes, as indicated in Figure 1.
www.cla-val.ch
CLA-VAL CV-Log-35
Quick Start Guide
cla-val@cla-val.ch
1 - CLOG35SE
D 07/23
Reduce your waste - Sort your rubbish

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CV-Log-35 and is the answer not in the manual?

Questions and answers

Summary of Contents for CLA-VAL CV-Log-35

  • Page 1 CV-Log-35 analog input while powered on could damage the product. Sensor connection If the CV-Log-35 sensors are not factory wired, start by connecting the desired sensors to the inputs of the product. Refer to the CVLog3500 wiring diagram for connection details.
  • Page 2 The default password is: CVLOG35_< last 4 digits of the series serial number > We strongly recommend changing the default password at first installation. Note: The network generated by the CV-Log-35 has no Internet access. Any error messages about this can be ignored. Enter the address http://192.168.4.1...
  • Page 3 Click on "SAVE SETTINGS" for the changes to be applied. Click on "SYNCHRONIZE NOW" to update your unit's clock immediately. This operation can only be performed if the CV-Log-35 is connected to the cellular network (option). If not, go directly to step 7.
  • Page 4 REGISTER ON LINK2VALVES (COMMUNICATION OPTION) Link2Valves is the CLA-VAL web platform (https://cla-val.ch) that allows the remote access to your CV-Log-35. A Link2Valves account is necessary. Please contact CLA-VAL to get one for free if you don't have it yet. Click on " Connectivity" menu.
  • Page 5 Associate the CV-Log-35 unit with your Link2Valves user account. First, enter the email address of your Link2Valves account. If you do not have one, please contact CLA-VAL to get one for free. Then click on "REGISTER NOW" and wait for the message "Success! ".

This manual is also suitable for:

D35

Table of Contents