HP 3325B Remote Operation Command Set
WIOAWJ; Modulation Source Amplitude Command
The MOAM command sets the amplitude of the modulation signal. Sending MOAM with no
value or units displays the current amplitude. Sending MOAM and units without any value
displays the current amplitude in the new units. MOAM? causes the instrument to output
the current amplitude.
^
Instrument Preset value: 0.1 Vpp
Command Availability
__MOAM_MOAM?
HP 3325B
Y^S
^
HP 3325A
No
No
Table 2-28. MOAM "value" Restrictions Given "units"
value range
units
Description
VO
Vpp
0.0-*12.0
MV
mVpp
0.0-* 12000.0
VR
Virus
0.0 -* 4.2
MR
mVrrns
0.0 ^ 4200.0
Table 2-29. MOAM? Response Format
Current Units
HEAD-on response
HEAD-off response
VO or MV
VRorMR
MOAM#####.######VO
MOAM#####.######VR
#####.######
#####.######
2-45