Table 5. Guidance and manufacturer's declaration - electromagnetic immunity
Test
* U
is the AC mains voltage prior to application of the test level.
T
Table 6. Guidance and manufacturer's declaration - electromagnetic immunity (conducted RF and
radiated RF)
Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
At 80 MHz and 800 MHz, the higher frequency range applies.
3
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
4
reflection from structures, objects, and people.
IEC 60601 Test Level (Actual Level)
IEC 60601 Test Level
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.7 GHz
Electromagnetic Environment - Guidance
Immunity Compliance
3
Level
3 Vrms
[V
=3]
1
3 Vrms
[E
=3]
1
12
Electromagnetic
Environment - Guidance
4
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the
Merlin@home
transmitter, including
cables, than the
recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance:
(80 MHz to 800 MHz)
(800 MHz to 2.7 GHz)
where
P
is the maximum
output power rating of
the transmitter in watts
(W) according to the
transmitter
manufacturer and
d
is
the recommended
separation distance in
meters (m).
Field strengths from
fixed RF transmitters, as
determined by an
electromagnetic site
survey*, should be less
than the compliance