TVOC MANUAL
4
TVOC Setup
IMPORTANT: if the 10ppm range is selected the TVOC will require a calibration using 10ppm Isobutylene
before use. (See the calibration section of this manual)
NOTE:
* TVOC is shipped with all selector pins fitted as standard
* TVOC only reads the selector pin settings when power is connected.
Always ensure power is disconnected from TVOC before changing
setting or carrying out maintenance.
* Never place selector pins on the programming port connector.
* Ensure static build up is discharged before touching components.
4.2 Duty Cycle
The TVOC has a variable duty cycle whereby the lamp is ON for 5 seconds and then OFF for a user set
period between 0 to 5 minutes in 1 second increments. The 4 – 20 mA output is updated at the end of each
cycle. As standard the TVOC duty cycle is 1 minute – the lamp is ON for 5 seconds and OFF for 55 seconds.
The lamp OFF time is fully adjustable by the user. Please see page 15of this manual for full instructions.
Page 10 of 28