Calibration

7.1 Calibration routine

Ion Science recommends calibrating TVOC after any maintenance or lamp cleaning is carried out and on a 3
monthly basis to ensure TVOC is working to specification.
NOTE: Please read this entire calibration procedure before attempting a calibration.
TVOC calibration has 4 stages:

1. Setting the ZERO

2. Setting the gas tolerance
3. Setting the SPAN
7.2 Setting the ZERO
1 –
Place the calibration magnet over the top right hand corner of the Ion Science logo to enter
calibration mode.
2 –
Remove the calibration magnet
3 –
Insert the calibration adaptor into the sensor cap, attach the carbon filter to the calibration
adaptor. The carbon filter should be attached to the hand aspirator.
4 –
Again place the calibration magnet over the logo.
displays a direct milivolt (mV) output from the PID sensor.
(Using clean gas via carbon filter)
(1 – 200 ppm Isobutylene on 0-100 ppm and 0-1000 ppm ranges)
(1 - 100 ppm Isobutylene on the 0-10ppm range)
(Using the SPAN gas)
The ZERO LED will illuminate and the GREEN STATUS LED will extinguish.
Calibration adaptor
Page 17 of 28
GREEN
STATUS LED
ZERO LED
SPAN LED
FAULT LED
The ZERO LED will flash, during this stage TVOC
Carbon filter
Ion Science Ltd
Hand
aspirator