Introduction Omega Bio-tek’s Mag-Bind® RXNPure Plus Kit allows rapid and reliable isolation of PCR products with high recovery rates. The system combines Omega Bio-tek’s proprietary chemistries with the reversible nucleic acid-binding properties of magnetic beads that selectively binds PCR amplicons 100 bp and larger and eliminates excess nucleotides, primers, and small, non-targeted amplification products, such as primer dimers.
Illustrated Protocol Measure the PCR Reaction Add Mag-Bind® RXNPure Plus and Mix Magnetize and Remove Supernatant Wash Twice with 70% Ethanol Elute DNA...
Mag-Bind® RXNPure Plus 96-well Plate Protocol Important: If automating this procedure on a liquid handler or a magnetic processor, please contact your Omega Bio-tek representative for instrument-specific instructions. Materials and Equipment to be Supplied by User: • Vortexer • Magnetic Separation Device (Recommended Cat# AlpAqua A001322) •...
Page 8
Mag-Bind® RXNPure Plus Add 1.8 volumes Mag-Bind® RXNPure Plus. PCR Reaction Volume (μL) Mag-Bind® RXNPure Plus (μL) Pipet up and down 5-10 times or vortex for 30 seconds. Let sit for 5 minutes at room temperature. Place the plate on a magnetic separation device to magnetize the Mag-Bind® RXNPure Plus.
Page 9
Mag-Bind® RXNPure Plus 13. Leave the plate on the magnetic separation device for 10-15 minutes to air dry the Mag-Bind® RXNPure Plus. Remove any residual liquid with a pipettor. Note: It is important to dry the Mag-Bind® RXNPure Plus before elution. Residual ethanol may interfere with downstream applications.
Mag-Bind® RXNPure Plus 384-well Plate Protocol Important: If automating this procedure on a liquid handler or a magnetic processor, please contact your Omega Bio-tek representative for instrument-specific instructions. Materials and Equipment to be Supplied by User: • Vortexer • Magnetic separation device for 384-well PCR plates •...
Page 11
Mag-Bind® RXNPure Plus Pipet up and down 5-10 times or vortex for 30 seconds. Let sit for 1 minute at room temperature. Place the plate on a magnetic separation device to magnetize the Mag-Bind® RXNPure Plus. Let sit at room temperature until the Mag-Bind® RXNPure Plus is completely cleared from solution.
Page 12
Mag-Bind® RXNPure Plus 16. Pipet up and down 20 times or vortex for 30 seconds. 17. Let sit for 2-3 minutes at room temperature. 18. Place the plate on a magnetic separation device to magnetize the Mag-Bind® RXNPure Plus. Let sit at room temperature until the Mag-Bind® RXNPure Plus is completely cleared from solution.
Troubleshooting Guide Please use this guide to troubleshoot any problems that may arise. For further assistance, please contact the technical support staff, toll free, at 800-832-8896. Possible Problems and Suggestions Problem Cause Solution Increase the number amplification cycles Low PCR product yield for PCR.
Need help?
Do you have a question about the Mag-Bind RXNPure Plus and is the answer not in the manual?
Questions and answers