Fluke 8024B Instruction Manual page 39

Hide thumbs Also See for 8024B:
Table of Contents

Advertisement

8024B
8024B
5.
Depress
the
AC/
DC
switch
to activate
the audible
tone.
6.
Momentarily
short the
test
leads
together
and
observe
that the
tone
sounds
coincident with
theTdown
arrow.
TheAup
arrow
may
or
may
not
be
displayed,
depending on
the
duration
of the
short.
4-37.
Level
Detector Test
4-38.
Use
the following
procedure
to verify
the
proper operation of
the
level
detector
function:
1.
Select the
ft
functions,
200
kft
range.
2.
Program
the pulse
generator
for
a single pulse that
is
greater
than 50
usee
wide
and 0
to
3V ±0.5%
in
amplitude.
3.
Connect
the pulse
generator
to
the
UUT:
+
to the
V/
ft/
S
terminal
and
-
to
the
COMMON
terminal.
4.
Cause
the pulse
generator
to
output
single
pulses
and
verify that the
Aup
arrow
appears
momentarily
in
the
LCD
of
the
UUT
for
each
single pulse.
5.
On
the
UUT
depress the
AC/
DC
switch
to
enable
the
audible
tone.
The
audible tone should be
on
continuously.
6.
Cause
the pulse
generator
to
output 500
ms
pulses
and
verify that the
A
up
arrow
appears
in
the
LCD
and
the
audible tone stops
for
each
pulse.
7.
Release
the
AC/
DC
switch
to
disable the
audible
tone.
The
down
arrow
should appear
in
the
UUT
display.
4-39.
BTTest
4-40.
Complete
the following
procedure
to verify that the
BT
indicator
appears
on
the
LCD
at
the correct battery
level,
and
that
the
accuracy
of
the
UTT
remains
unaffected
at
this
battery voltage
level:
1.
Connect
the
equipment
as
shown
in
Figure
4-5.
2.
Set the
UUT
switches
to the
following
positions:
200
mV
in
AC/
DC
DC
(out)
3.
Set the
DMM
controls
to
the
following
positions:
in
DC
Figure
4-5.
BTTest
4.
Program
the
DMM
Calibrator
for a
UUT
input of 190.0
mV
dc.
5.
Adjust
the variable
power
supply
until
the
BT
indicator
appears
in
the
UUT
display.
6.
Verify
that the
DMM
display
is
between +6.5 and
+7.5V
dc.
7.
Decrease
the
output
of
the variable
supply
until
the
DMM
displays
+6.0V.
8.
Verify that the
UUT
display
is
between
189.8
and
190.2
mV
dc.
9.
Program
the
DMM
Calibrator
for
an
input of
0V
dc.
10.
On
the
UUT,
depress the
ft/
S
function switch
and
2 kft switch.
1
1.
Adjust
the variable
power
supply
until
the
DMM
displays
+I0.0V
dc.
12.
Program
the
DMM
Calibrator
for
a
UUT
input of
1
kft.
13.
Verify
that the
UUT
display
is.
between
0.998
and
1.002.
14.
Adjust
the variable
power
supply
until
the
DMM
displays
+6V
dc.
15.
Verify that the
UUT
display
is
between
0.998
and
3.002.
4-12