Alarm 109:
Sample line signal error
Alarm no.
Alarm displayed on front panel
Front-panel LED
Displayed by SPECTLAND2
Contact output
Level
Auto recovery
DATA1
DATA2
Location detected
Method of detection
Processing
User action
Notes
Alarm 110:
Intrtferogram data failure
Alarm no.
Alarm displayed on front panel
Front-panel LED
Displayed by SPECTLAND2
Contact output
Level
Auto recovery
DATA1
DATA2
DATA3
DATA4
Location detected
Method of detection
Processing
User action
109
STRM c (c: channel no.)
Alarm LED
Sample line signal error
NR800L fail alarm contact
Intermediate priority (M)
Yes
Channel no.
0: Stream signal obtained from DCS
1: Stream signal obtained from module A
2: Stream signal obtained from module B
CPU (stream mode switching task)
Channel no. and stream no. are related. Stream signal can t be
acquired.
Stream switching command invalid
Check I/O unit or DCS that should be providing stream no.
Channel no. and stream no. correspondence depend on channel mode,
no. of channels, and whether or not channel switching is supposed to
occur. Error occurs if stream no. is invalid. E.g. if communications with
DCS fails and old channel no. info. remains.
110
Alarm LED
Intrtferogram data failure
NR800L fail alarm contact
Intermediate priority (M)
Yes
0
0
0
0
DSP Card (DSP)
Scan direction changes before data collection
None
Contact our service representative in case frequently happened.