Temperature Correction Table

For Refractometers Calibrated At 20°C
As an alternative to recalibration of the refractometer due to ambient temperature changes, the following tempera-
ture correction table can be used. Ensuring that both the distilled water and the ambient temperature are both at
exactly 20°C, follow steps 2-7 under the CALIBRATION section on the preceding page. Once this is accomplished, if
ambient temperature changes, simply apply the temperature correction values listed below. For example, a reading
of Brix 40.0% at 25°C is corrected to Brix 40.4%; a reading of Brix 65.0% at 10°C is corrected to Brix 64.25%.

Specifications

Range:
Brix 0.0-32.0%
Resolution:
0.2%
±0.2%
Accuracy:
40 x 40 x 165mm (1.6 x 1.6 x 6.5")
Dimensions:
Weight:
240g (8.5 oz.)
Supplied With:
Vinyl Carrying Case (1), Plastic Transfer Pipet (1)
Printed in China
(l
589nm)
=
Source: ICUMSA, 1974
Warranty information and
registration form can be found at:
www.veegee.com/service_support
VGMNL010320-43001
Operation Manual
Model BX-1
Brix 0-32%
Cat. No. 43001

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BX-1 and is the answer not in the manual?

Questions and answers

Summary of Contents for VEE GEE BX-1

  • Page 1 For example, a reading of Brix 40.0% at 25°C is corrected to Brix 40.4%; a reading of Brix 65.0% at 10°C is corrected to Brix 64.25%. Operation Manual Model BX-1 Source: ICUMSA, 1974 Specifications Range: Brix 0.0-32.0%...
  • Page 2 Calibration should be conducted at the start of each day or when any Thank you for purchasing this VEE GEE Refractometer. With the user in mind, VEE GEE Refractometers are built shifts in ambient temperature occur. If recalibration is impractical, refer from modern designs and, with proper care, this instrument should provide many years of reliable performance.

Table of Contents