Temperature Correction Table
For Refractometers Calibrated At 20°C
(l
As an alternative to recalibration of the refractometer due to ambient temperature changes, the following temperature
correction table can be used. Ensuring that both the distilled water and the ambient temperature are both at exactly 20°C,
follow steps 2-7 under the CALIBRATION section on the preceding page. Once this is accomplished, if ambient temperature
changes, simply apply the temperature correction values listed below. For example, a reading of Brix 40.0% at 25°C is
corrected to Brix 40.4%; a reading of Brix 65.0% at 10°C is corrected to Brix 64.25%.
Specifications
Range:
Brix 45.0-82.0%
Resolution:
0.2%
Accuracy:
±0.2%
Dimensions:
40 x 40 x 140mm (1.6 x 1.6 x 5.5")
Weight:
230g (8.1 oz.)
Supplied With:
Vinyl Carrying Case (1), Plastic Transfer Pipet (1)
Brix 60.0%, ¼ oz. Bottle (cat. no. 45000-OS025)
Optional Oil Standards:
Printed in China
=
589nm)
Source: ICUMSA, 1974
Warranty information and
registration form can be found at:
www.veegee.com/service_support
VGMNL010320-43007
Operation Manual
Model BX-4
Brix 45-82%
Cat. No. 43007

Advertisement

loading

Summary of Contents for VEE GEE BX-4

  • Page 1 For example, a reading of Brix 40.0% at 25°C is corrected to Brix 40.4%; a reading of Brix 65.0% at 10°C is corrected to Brix 64.25%. Operation Manual Model BX-4 Source: ICUMSA, 1974 Specifications Range: Brix 45.0-82.0%...
  • Page 2 Calibration should be conducted at the start of each day or when any shifts in Thank you for purchasing this VEE GEE Refractometer. With the user in mind, VEE GEE Refractometers are built from modern ambient temperature occur. If recalibration is impractical, refer to the directions designs and, with proper care, this instrument should provide many years of reliable performance.