Alarm and Alert Screen Descriptions

Low Purity Alarm
(technical)
(high priority)
Solutions
Low Battery Alarm
(technical)
(high priority)
Solutions
The LCD screen will be RED in Alarm Mode
When oxygen levels drop below 72% for 60 seconds,
the screen will flash red, sound a 10-beep sequence
repeated after 2.5 seconds of silence and the text "Low
Purity" will flash on the screen. The alarm will continue
to sound until the power
momentarily to silence the audio; or the power button is
held to power off the unit; or 15 seconds has expired at
which time the device will power off.
Clean or replace side vents.
Change to another source of oxygen and contact your
authorized dealer.
When battery power is depleted, the pump shuts off, the
screen will flash red at 2Hz, sound a 10-beep sequence
repeated after 2.5 seconds of silence and the text "Low
Battery" will flash on the screen. The alarm will continue
to sound until the power
momentarily to silence the audio; or the power button is
held to power off the unit; or 15 seconds has expired at
which time the device will power off.
Plug into an external power source. Replace depleted
battery with a charged battery.
Alarms
button is pressed
button is pressed