Oxlife Freedom
®
Provider Orientation
Guide
®
O2-CONCEPTS.COM | 877-867-4008
800-1063 REV A

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Oxlife Freedom and is the answer not in the manual?

Questions and answers

Summary of Contents for O2 Concepts Oxlife Freedom

  • Page 1 Oxlife Freedom ® Provider Orientation Guide ® O2-CONCEPTS.COM | 877-867-4008 800-1063 REV A...
  • Page 2: Feature Identification

    Introduction The Oxlife Freedom is a portable oxygen concentrator used on a prescriptive basis by patients requiring ® supplemental oxygen to be treated in a home, office, or vehicle/mobile environment. The Freedom delivers 87%-95% pure oxygen to the patient through a standard single lumen nasal cannula.
  • Page 3: Device Settings

    Device Settings Pulse dose settings 1-5 deliver the following volumes (mL) at 20 breaths per minute. Setting Setting Setting Setting Setting 16.0 24.0 32.0 40.0 External Power To use the AC power supply, connect the power cord to the AC power converter (brick).
  • Page 4 Any battery with a capacity of less than 70% or less than 4,480mAh should be replaced. Battery capacity readings can be accessed using the device serial number on the O2 Concepts’ DNA portal at: www.o2dnaportal.com.
  • Page 5: Device Monitoring

    If the Freedom has not connected within 15 minutes, power off and remove all power sources (batteries, AC/DC power supply). After 30 seconds, reconnect the power supply and power the device on to repeat the process. Device Monitoring Access O2 Concepts’ DNA portal at: www.o2dnaportal.com Enter user name and password. Page 4 of 7...
  • Page 6: Maintenance

    Provider Screen To access the Provider Screen, press the Menu (M) button. Then press the Plus (+) button or Minus (-) button to scroll to the Provider Screen icon. To enter and exit the Provider Screen, press the Menu (M) button again.
  • Page 7: Carry Bag

    Device Care and Cleaning Cleaning Unplug your Freedom before cleaning. Do not submerge your device or any accessories or allow water to enter the device. Clean the exterior of your device using a soft cloth dampened with a mild detergent and water and wipe dry.
  • Page 8: Device Identification

    Battery power is depleted, device will shut down Invalid Flow Pulse bolus volume is out of valid range Unauthorized Battery A non-O2 Concepts battery is detected and no other power source is available High External Power Voltage to the device is above 26 volts...
  • Page 9 ® O2-CONCEPTS.COM | 877-867-4008...