10. Troubleshooting and service

Only a technician authorized by Naggura is allowed to perform any kind of repairs on the bed.
None of the parts inside the device may be replaced by the user. If your technical staff is allowed by Naggura to repair the
bed, naggura can provide the necessary instructions and spare parts lists.
 
Please contact your local Naggura distributor for more information.
If you need to reach the electrical main components
of the table to perform repairing tasks, the table must
be leant laterally to let it resting on the floor. To do so,
the operator has to install a protection kit to prevent
the Space Ball mechanisms from being damaged. If
not, the sphere metal mechanisms will bend.
First, install the set of 2 L-shaped plates as it is shown
in the picture to turn the bed to one side without da-
maging the Space Ball perimeter switch.
Take the decorative panel in the base out and screw
the L-shaped plates in the same threaded holes.
Place a washer between the table frame and the
L-shaped plate and screw it tightly with the Allen key
provided. The mounting tools, screws and washers
are provided by Naggura.
When both L-shaped plates are installed at each
side, the table can be leant to the side safely. 2 peo-
ple needed to tilt the table carefully.
In order to protect the upholstery, we recommend you
to put a towel or a protective tarpaulin on the ground.
Squeaking noise when moving the table up & down (mechanical noise)
• The axes are adjusted and greased from factory but with the time and use, they may need some little adjustement. Remove
the grey plugs of the 2 connecting rods (4 at each side, so 8 in total). Loosen the axes a litlle bit by unscrewing them (approx. 2
turns). Apply a bit of neutral grease and plug them again.
17