Model 5328A
General Information
Table 1-3. Specifications (Continued)
GENERAL
Display:
8 digit (9 with Opt. 030) LED display
Blanking:
Suppresses display of unwanted zeros to left of
most significant digit
Storage:
Holds reading between samples; can be overridden
by rear panel switch.
Sample rate: Variable from less than 2 ms between measure-
ments to HOLD which holds display indefinitely.
Gate output:
Rear panel output; TTL levels; high when
counter gate open
Timebase output:
Rear panel output; TTL
levels
Check signal: With function switch in CHECK, counter
should display 10 MHz
+1 count. (With opt. 040, place
function switch in Freq A and universal module
in CHECK
(CHK) — counter should display 100 MHz
+1 count.)
Timebase:
standard crystal
Aging rate:
<3 x 10~'/month
Temperature:
<2.5 x 1076 0° to 50°C
Line voitage:
<1 x 10-7 for 10% change
Opt. 010 over oscillator
Aging rate:
<5 x 10-'°/day after 24-hour warm-up
Short term:
<1 x 10-'° rms/sec
Temperature:
<7 x 107° 0° to 50°C
Line voltage: <+5 x 10-8 for 10% variation
Warm-up:
<+5 x 10-9 in 20 min.
Ext. Freq. Std. Input: 30 kHz to 10 MHz signal of amplitude
>1.0 V rms into 1 kQ2. Maximum input: 5 V p-p. For
opt. 040 only, the following constraints apply: ext. freq.
std. must be 10 MHz for Period Avg., T.I. Avg., Period
(N=1), and T.1. (N=1).
Trigger Lights: Light is ON when input is above trigger level;
OFF when input is below trigger level; BLINKING when
channel is triggering. Operative over frequency range
0—100 MHz.
Marker Outputs:
inverted channel A and channel B Schmidt
trigger outputs available on front panel; 0 to -100 mV
levels into 502 ;<20 ns delay. (With Opt. 040, inverted
channel A Schmidt trigger and T.I. A*B
marker outputs
(0 to -50 mV) available on front panel—T.!. A—*B
is high
during the time interval measured
by the counter). Outputs
protected from inadvertently applied voltage to + 5 Vdc.
ARM:
Rear panei switch turns arming ON or OFF. With
arming ON, the measurement is armed by an input other
than the input involved in the measurement. The following
are armed by an event at B: FREQ A, PERIOD
A, PERIOD
AVG A, FREQ C, DVM, C/A; the following are armed by an
event at C: T.l. AB,
T.l. AVG A+B, EVENTS C, AB,
B/A.
Operating Temperature:
0° to 50°C
Power Requirements:
100/120/220/240 V rms, +5%, -10%
(switch selectable), 48—66
Hz; 150 VA max.
Accessories Furnished:
Power cord, 200 cm
(7% ft.)
Weight: Net 8.5 kg (18 lb, 12 0z). Shipping 11.2 kg (24 Ib. 12.02)
Dimensions:
ees
(18 18) ——»|
NOTE
|
DIMENSIONS
ARE
IN MILLIMETERS
}
:
f
AND
(iNCHES)
HANDLES
ARE
REMOVABLE
ae a
le
=}
«422
(16-5'8)
)
89
(31/2)
T=
Jil 7]
|
438
:
(171-4)
Available Accessories
HP Rack Mount
Ears (2 each required)
(for use with handles)
(for use without handles)