MDT CASTLE 5520 Service Manual page 21

Warming cabinets
Table of Contents

Advertisement

5520, 5525, & 5530 WARMING CABINETS
3. Trouble Analysis
Red Warning Lit
Condition #1
Red Warning code,
OF
in temperature
display.
Condition #2
Red Warning code,
temperature
indi-
cation in display.
Has
chamber
set
Is red code still on?
Replace PC board
No
Check
wiring
and
connections
for
open
circuits.
Re-
place
temperature
sensor.
Refer to Operator's
Yes
temperature
been
Manual for temper-
decreased accident-
ature reduction tech-
ally or intentionally
nigue or reset temp-
erature to intended
set point
Check PS and PCB
calibration.
Read-
just if necessary.
1
Calibration good or
adjustable.
Yes
Recheck
unit
for
proper operation.
No
Unit
operation
good
Replace PC board
Yes
Unit good
16
30926

Advertisement

Table of Contents
loading

This manual is also suitable for:

Castle 5525Castle 5530