TE-4602
Tinytag Plus LAN Data Logger with Two Temperature & Relative Humidity Probes
(-25 to +85°C/0 to 100% RH) x 2
• Ethernet connectivity (communicates across a LAN
• Two temperature and relative humidity probes
• User programmable audible & visual alarm
• Alarm output, to trigger external device
• Power over Ethernet (PoE) with battery back-up
• Waterproof case
• Tinytag Connect compatible
The TE-4602 is a temperature and relative humidity data logger that plugs
directly into a network point and communicates across a LAN.
The logger can be managed as a standalone logger, in conjunction with the
standard version of Tinytag Explorer, or as part of a network of loggers, using
Tinytag Connect software.
The data logger measures temperature and relative humidity using two
probes.
The logger records data at user-defined logging intervals and can be
programmed to give alarm alerts, both audible and visual, in the event of a
problem. The logger also has an alarm output that changes state when alarms
are active.
Standalone
When used as a standalone device, the logger connects to a standard version
of Tinytag Explorer across a LAN and is downloaded and configured directly.
Tinytag Connect
When used as part of a Tinytag Connect system, the logger transmits data to
a Windows service that runs the system, called the Connect Gateway. This
data is then viewed, and loggers configured, using the Connect version of
Tinytag Explorer.
Multiple users can access the Connect system simultaneously.
If a logger cannot communicate with the system for any reason, data will be
recorded locally until communications are restored, ensuring no data is lost.
The Connect system can also send e-mail warnings when alarm limits are
exceeded.
Working With the Data
Using Tinytag Explorer, data recorded by the logger can be viewed as a graph
or as a table of readings. There are also summary views containing
information about the recording and a daily min/max view.
Data from multiple devices recording at the same time can be combined into a
single graph.
Graphical data can be exported from the software as an image, for use in
report writing or presentations, or as a data table, for further analysis in third-
party spreadsheet and data base programs.
www.tinytag.info
sales@tinytag.info
Issue 2 (27th October 2023) E&OE

Advertisement

loading

Summary of Contents for Gemini Tinytag Plus TE-4602

  • Page 1 TE-4602 Tinytag Plus LAN Data Logger with Two Temperature & Relative Humidity Probes (-25 to +85°C/0 to 100% RH) x 2 The TE-4602 is a temperature and relative humidity data logger that plugs directly into a network point and communicates across a LAN. The logger can be managed as a standalone logger, in conjunction with the standard version of Tinytag Explorer, or as part of a network of loggers, using Tinytag Connect software.
  • Page 2: Data Logging

    TE-4602 Tinytag Plus LAN Data Logger with Two Temperature & Relative Humidity Probes (-25 to +85°C/0 to 100% RH) x 2 Data Logging Physical Specification (Logger) This data logger will turn on and start recording as soon as it is plugged into a Power over Ethernet (PoE) supply (it can also be turned on before that by pressing and holding the logger's Mute button briefly) .
  • Page 3 TE-4602 Tinytag Plus LAN Data Logger with Two Temperature & Relative Humidity Probes (-25 to +85°C/0 to 100% RH) x 2 Physical Specification (Probe) Mounting Instructions (Logger) Suggested positioning for the RH probe if using the supplied clip to mount it on the 2 x PBRF-8002 temperature and relative humidity probes and CAB-0034 logger.
  • Page 4: Power Information

    TE-4602 Tinytag Plus LAN Data Logger with Two Temperature & Relative Humidity Probes (-25 to +85°C/0 to 100% RH) x 2 Alarms Power Information LAN data loggers can be programmed with alarm limits to warn when a This logger is a PoE (Power over Ethernet) device that is powered from its Ethernet connection (a PoE injector is supplied).
  • Page 5: Warranty

    SWCD-0080 Tinytag Explorer Connect Approvals Probes Gemini Data Loggers (UK) Limited hereby declares that this data logger is in This logger uses 2 x PBRF-8002 probes, which are fitted to the loggers using compliance with the essential requirements and other relevant provisions of CAB-0034 extension cables for Plus &...