Page 1
Model 1404-02ADB FATIGUE RATED LOW PROFILE LOAD CELL, DUAL BRIDGE Installation and Operating Manual For assistance with the operation of this product,contact: PCB Load & Torque, Inc. Toll-free: 866-684-7107 24-hour SensorLine™: 716-684-0001 Fax: 248-888-8266 E-mail: LTInfo@pcbloadtorque.com Web: www.pcbLoadTorque.com...
Page 2
“Lifetime Warranty Plus” on all Platinum Stock Products number, which we use to reference any information you sold by PCB and through its limited warranties on all other have already provided and expedite the repair process. PCB Stock, Standard and Special products. Due to the...
Page 3
CAUTION Refers to hazards that could damage the instrument. NOTE Indicates tips, recommendations and important information. The notes simplify processes and contain additional information on particular operating steps. The following symbols may be found on the equipment described in this manual: This symbol on the unit indicates that high voltage may be present.
3.1 Dimensions 1.0 INTRODUCTION The following figure and table give the general outline Dual bridge fatigue-rated load cells manufactured by PCB dimensions of the dual bridge fatigue-rated load cells. Load and Torque, Inc. are guaranteed to last 100 million fully reversed cycles (full tension through full compression).
25%, 50% and finally 100% of the NOTE: Pre-tension studs include the threaded stud and the required torque being applied. jam nut. PCB Load & Torque, Inc. Toll-Free in USA 866-684-7107 716-684-0001 www.pcbloadtorque.com...
#4-40 x 3/8” Socket Head Cap Screw Figure 7 - Remove the #4-40 x 1/4” Figure 8 - Slide the Connector Button Head Cap Screws Protector (54823-02) Behind the Connector PCB Load & Torque, Inc. Toll-Free in USA 866-684-7107 716-684-0001 www.pcbloadtorque.com...
+ Signal Green 6.0 CALIBRATION + Excitation Red Every dual bridge fatigue-rated load cell purchased from PCB Figure 11 - Western Regional Strain Gage Committee Wiring Code Load & Torque, Inc. has been fully calibrated per ISO/IEC 17025 procedures, and meets all published specifications.
14. The maximum shear stress measured by the bonded strain to minimum load. gages is less than 40% of the endurance limit shear stress for alloy steel and 48% for aluminum. PCB Load & Torque, Inc. Toll-Free in USA 866-684-7107 716-684-0001 www.pcbloadtorque.com...
–P on them. Never use a pressure washer on the load cells. connector leads. PCB Load & Torque, Inc. Toll-Free in USA 866-684-7107 716-684-0001 www.pcbloadtorque.com...
12.0 CALIBRATION / REPAIR SERVICES conductors for short circuits. Check isolation resistance, load cell flexure to PCB Load & Torque offers calibration and repair services. conductors. The PCB Calibration Laboratory in Farmington Hills, Michigan is A2LA Accredited per ISO/IEC 17025.
(SEB) calculations and entries abilities and traceability statements. If an initial evaluation shows that a transducer requires repair, PCB will provide the customer with an estimate prior to taking any corrective action. 12.1 RMA / Purchase Order Please request a return material authorization (RMA) before sending a load cell back to the factory for any reason.
Page 15
0.03 mm Electrical Connector PT02E-10-6P (2) PT02E-10-6P (2) PCB Load & Torque, Inc. All specifications are at room temperature unless otherwise specified. In the interest of constant product improvement, we reserve the right to change specifications without notice. 24350 Indoplex Circle...
Page 16
PCB Load & Torque Inc. claims proprietary rights in REVISIONS the information disclosed hereon. Neither it nor any reproduction thereof will be disclosed to others without DESCRIPTION the written consent of PCB Load and Torque Inc. CONNECTOR WAS PT02A - 06.06.16, PTE 45560 CONNECTOR "B"...
Need help?
Do you have a question about the 1404-02ADB and is the answer not in the manual?
Questions and answers