Page 4
“Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Keysight Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987).
Page 5
Use the following guidelines with cables: • Never coil the cable too tightly. • Never allow the cable to hang down from an instrument test port, or a storage fixture, or while holding it. • Never bend the cable an excessive number of times. 85133E/F/H...
Page 6
• Go to http://www.keysight.com. • Enter the document’s part number (located on the title page) in the Search box. • Click Search. • Click on the hyperlink for the document. • Click the printer icon located in the tool bar. 85133E/F/H...
The Keysight 85133E/F/H flexible cable sets are designed for use with Keysight network analyzers that have male NMD-2.4 mm test ports. 85133E The 85133E set contains a single cable. Use it when a DUT is connected, with the appropriate adapter, directly to the test set test port. See Figure 1-1.
When you send the cable set to Keysight, include the following information: • your company name and address • the name of a technical contact person within your company, and the person’s complete phone number • the model number and serial number of the cable set 85133E/F/H...
(you can order a single cable rather than the entire cable set), and for recommended items not included with the cable set. Refer to “Contacting Keysight” on page 4-3 information about communicating with Keysight. 85133E/F/H...
Up to 80% at 30C Storage Up to 95% at 40C Electrical Specifications Table 2-2 Electrical Specifications Cable Return Loss Insertion Frequency Range (dB) (GHz) Loss (dB) 85133E 1.44 0.446 DC to 50 0.017F + 0.456F 85133F 0.450 0.013F+ 0.300F 85133H a.
Length Electrical Phase Stability Magnitude and Recommend Length With a 90 Phase ed Bend Radius Bend Stability 85133E 97.2 38.25 1.125 3.690 <0.15 dB Change <0.07 dB Change 6.352 <0.17 (f) + 0.5 <0.09 (f) + 0.5 ...
Page 16
Connector NMD-2.4 mm -f- -0.000 to -0.056 -0.0000 to -0.0022 0.0000 0.0000 NMD-2.4 mm -m- -0.0025 to -0.0254 -0.0001 to -0.001 PSC-2.4 mm -f- -0.0025 to -0.0254 -0.0001 to -0.001 a. Center conductor shoulder behind outer conductor mating plane. 85133E/F/H...
2. While wearing a grounded wrist strap, grasp the outer shell of the cable connector. 3. Connect the other end of the cable to the test port. 4. Remove the short from the cable. Refer to Chapter 4 for part numbers and instructions for ordering ESD protection devices. Figure 3-1. ESD Protection Setup 85133E/F/H...
If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again. Cables with damaged connectors should be repaired or discarded. Determine the cause of damage before connecting a new, undamaged connector in the same configuration. 85133E/F/H...
The precision slotless connector was developed to eliminate this problem. The PSC has a center conductor with a solid cylindrical shell, the outside diameter of which does not change when mated. Instead, the center conductor has an internal contact that flexes to accept the male pin. 85133E/F/H...
Apply a small amount of isopropyl alcohol to a lint-free cleaning swab. b. Clean the connector threads. c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean, low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it. 85133E/F/H...
Page 22
Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean, low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it. 4. Reinspect Inspect the connector again to make sure that no particles or residue are present. 85133E/F/H...
Use, Maintenance, and Care of the Cables Gaging Connectors Gaging Connectors The gages available from Keysight Technologies are intended for preventive maintenance and troubleshooting purposes only. (See Table 4-1 on page 4-4 for part number information.) They are effective in detecting excessive center conductor protrusion or recession, and conductor damage on test cables and other accessories, DUTs, and calibration kit devices.Do not use the...
After each measurement, rotate the gage a quarter-turn to reduce measurement variations that result from the gage or the connector face not being exactly perpendicular to the center axis. e. Compare the average reading with the observed pin depth limits in Table 2-3 on page 2-4. 85133E/F/H...
Page 25
Use, Maintenance, and Care of the Cables Gaging Connectors Figure 3-2 Gaging 2.4 mm Connectors 85133E/F/H...
Refer to Table 4-1 on page 4-4 replacement part number and ordering information. Table 3-1 Torque Wrench Information Connector Type Torque Setting Torque Tolerance 2.4 mm 90 N-cm (8 in-lb) 9.0 N-cm (0.8 in-lb) 3-10 85133E/F/H...
Page 27
3. Apply downward force perpendicular to the wrench handle. This applies torque to the connection through the wrench. Do not hold the wrench so tightly that you push the handle straight down along its length rather than pivoting it, otherwise you apply an unknown amount of torque. 85133E/F/H 3-11...
• Do not set cable connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface. 3-12 85133E/F/H...
After you perform a calibration, move the cables as little as possible. Every time NOTE you bend a cable, the phase changes slightly. 85133E/F/H 3-13...
Page 30
Use, Maintenance, and Care of the Cables Handling and Storage 3-14 85133E/F/H...
Ordering Replaceable Parts Ordering Replaceable Parts Table 4-1 on page 4-4 lists the replacement part numbers for items included in the 85133E/F/H cable sets. To order a listed part, note the description, the part number, and the quantity desired. Telephone or send your order to Keysight Technologies (see “Contacting Keysight”...
Replaceable Parts Returning a Cable or Cable Set to Keysight Returning a Cable or Cable Set to Keysight If your cable set requires service, contact Keysight Technologies (see below). Include the following information: • your company name and address • a technical contact person within your company, and the person's complete phone number •...
Page 34
Replaceable Parts Returning a Cable or Cable Set to Keysight Table 4-1 Replaceable Parts Keysight Part Description Cables 85133E Single Cable: NMD-2.4 mm –f– to PSC-2.4 mm –f– 85133-60015 85133F Cables: NMD-2.4 mm –f– to NMD-2.4 mm –m– 85133-60017 NMD-2.4 mm –f– to PSC-2.4 mm –f–...
Page 35
00901-60001 2.4 mm 50 fixed termination -f- 00901-60002 2.4 mm offset short -m- 85056-60001 2.4 mm offset short -f- 85056-60002 2.4 mm airline (5.0 cm) 85057-60001 a. Refer to “Clarifying the Terminology of a Connector Interface” on page 1-3. 85133E/F/H...
Page 36
Replaceable Parts Returning a Cable or Cable Set to Keysight 85133E/F/H...
Proper connector care and connection techniques are critical for accurate, repeatable measurements and for extending the life of your devices. Prior to making connections to the network analyzer, carefully review the connector care information provided with your product. See the following table for quick reference tips about connector care. 85133E/F/H...
Page 39
Twist or screw any connection Use a torque wrench for final connect Tighten past torque wrench “break” point a. Use isopropyl alcohol in a well-ventilated area, allowing adequate time for moist alcohol to evaporate and fumes to disperse prior to energizing equipment. 85133E/F/H...
Page 40
Connector Care Quick Reference Principles of Microwave Connector Care 85133E/F/H...
Page 41
NMD type zeroing PSC type gage master quick reference part numbers gaging slotless connectors threads when to do type gender, connector DUT end test set end visual inspection handling 3-12 wear humidity connector terminology contacting Keysight Index-1 85133E/F/H...
Page 43
Index warranty, documentation wrench open-end 3-11 3-12 part numbers proper positioning of 3-11 torque 3-10 3-11 precautions for use of 3-11 proper use of 3-11 zeroing connector gage Index-3 85133E/F/H...
Need help?
Do you have a question about the 85133E and is the answer not in the manual?