Test data
M Pressure characteristics<Reference only>
Condition : Power supply : 50Hz Local voltage
AT : 35°C
4.0
3.5
3.0
2.5
2.0
15
1.0
0.5
peak 3.19MPa(32.5kg/cm°abs)
High pressure side(Pd)
2.02MPa(20.6kg/cm? abs)
Low pressure side(Ps)
0.15MPa(1.5kg/cm? abs)
1
1
i
1
L
1
Om
10m
20m
30m
40m
50m
th
2h
3h
4h
Sh
6h
Time scale
Note:In measuring gas pressure.
+ Make sure that in high pressure side. Residual refrigerant in measuring hose were collected to
the refrigeration circuit.
+ Make sure to evacuate the measuring hose and then open the refrigeration circuit
prior to the
measurement in low pressure side.
Condition : Power supply : 60Hz Local voltage
AT : 35°
4.0
3.5--
3.0
25
2.0
1.5
1.0
0.5
peak 3.3MPa(33.8kg/cm"abs)
High pressure side(Pd)
2.0MPa(20.8ka/cm? abs)
Low pressure side(Ps)
0.11MPa(1.1ka/cm? abs)
L
1
0m
10m
20m
30m
40m
50m
1h
2h
3h
4h
5h
6h
Time scale