Controls And Indicators - HP 59501A Operating And Service Manual

Hp-ib isolated d/a power supply programmer
Hide thumbs Also See for 59501A:
Table of Contents

Advertisement

SECTION IH
OPERATING INSTRUCTIONS
3-1
INTRODUCTION
3-2
This section contains ocerating instructions and
programming information for the 59501A
Isolated D/A,
Power Supply programmer.
Since the 59501 A can be used
as a power supply programmer or as D/A converter (low
level dc signal source), operating instructions and program-
ming examples are provided to cover both applications.
Additional information regarding use of the 595014
is
provided in HP-IB/Power Supply Application Guide, HP
Part No. 5952-3990,
3-3
PRE-OPERATIONAL CONSIDERATIONS
3-4
Before connecting the 59501 A to an ac power
source, ensure that the proper operating voltage (100, 120,
220, or 240Vac) has been selected and the proper fuse is
installed (see paragraph 2-27).
Also, check that the address
switches are set correctly, the UNIPOLAR/BIPOLAR
mode
switch is set to the desired position, and all HP-IB system
cabling is installed (see paragraphs 2-20 through 2-26).
Connections between the 59501A and the user's device
(e. g. HP Power Supply) are described in subsequent para-
graphs.
3-5
CONTROLS AND INDICATORS
3-6
Front Panel
3-7
The 59501A
front panel contains a power on indi-
cator, a status indicator and four controls (screwdriver ad-
justments) as shown in Figure 3-1.
3-8
The power on (ON) indicator () (Figure 3-1)
lights and the 59501A's internal supply voltages (+15V,
t15V, V aa! are present when the line cord is plugged-in.
3-9
The LISTENING indicator (2) lights when the
595014
is addressed to "listen" and remains iighted until
an 'unlisten'' command or the interface clear signal is
received from the controller.
When lighted, it indicates that
the 59501
is enabled to process data words received on
the HP-IB.
When it is not lighted, it indicates that the
59501
is inhibited from processing data words.
3-10
The front panel screwdriver adjustments allow fast
and easy calibration of D/A and power supply outputs.
Each adjustment is described below.
3-11
The ZERO ADJUST
@) allows the 59501A out-
put to be adjusted to OV
250 millivolts when it is program-
med to zero output,
It can also be used to correct for small
offsets in power supply response to programmed inputs when
the 59501A
is used as a power supply programmer.
3-12
The D/A FULL SCALE ADJUST
(â) allows the
59501A output to be adjusted to maximum
(£556) when it
is programmed to maximum.
For example, if the 59501A
is programmed to its maximum unipolar output in the high
range, the D/A FULL SCALE ADJUST is normally used to
set the output to 9.99V.
However, it can also be used to
set the output between 9.49V and 10.49V (i. e. 9.99V 45%).
The D/A FULL SCALE ADJUST is used when the 59501A
is employed as a D/A (see paragraphs 3-75 through 3-82).
When the 59501A is used as a power supply programmer,
the POWER SUPPLY
FULL SCALE ADJUST controls are
utìlized to set the power supply's output.
set the maximum output desired from the power supply
when the 59501A is programmed to its maximum value.
These adjustments can be used to calibrate the output of a
power supply to its maximum rated output or to any lower
value.
For example, a 40Vdc power supply's output could
be set to 20V when the 59501A
is programmed to its maxi-
mum value.
In this case, all 999 programming steps are
utilized in programming the supply from 0 to 20V with a
resolution of 20mV (approx.).
Calibration of power supply
outputs is described in paragraph 3-53 through 3-59.
ZERO
KBJUST
=
(B
Figure 3-1. 59501A Front Panel

Advertisement

Table of Contents
loading

Table of Contents