INSTRUCTIONS
REASON FOR CHANGE
INSERT THIS SHEET INTO THE MANUAL
LOW PRESSURE GAS MODULE
DESCRIBED BELOW.
EQUIPMENT:
Resuscitaire Radiant Warmer, Model RW82
MANUAL TITLE:
Service Manual
MANUAL CAT.
NO.:
82 990 07
ADDENDUM
INFORMATION
MAKE THE FOLLOWING CHANGES WHEN A LOW PRESSURE GAS MODULE IS PRESENT IN THE
RESUSCITAIRE RADIANT WARMER:
TABLE 3.1A SPECIFICATIONS
Page 3-3:
Change cylinder pressure from 2900 psi max (199.8 bar) to 3000 psi max (207 bar)
TABLE 3.1b SPECIFICATIONS
Page 3-5:
Change cylinder pressure from 199.8 bar to 207 bar.
SECTION 3 RESERVE GAS SUPPLY PAGE 3-18. ADD THE FOWLING STATEMENT:
The regulator for United Kingdom and Canadian units is preset to 45 psi when the cylinder pressure is 1500
psi.
TABLE 6.10 RESUSCITAIRE RADIANT WARMER О PIPELINE (DISS/NIST) WITH O2 CYLINDER (PIN-
INDEX) GAS SUPPLY ASSEMBLY, PARTS LIST ADDS THE FOLLOWING:
Gas Supply, Oz Pipeline (NIST) O2 with Cylinder (Pin-Index) Low Pressure - Part Number 81501 94.
Add to Item 23 regulator, 45 psi W/press relief - Part number 81 501 77.
TABLE 6.12 RESUSCITAIRE RADIANT WARMER
OyAIR
PIPELINE (DISS/NIST) WITH
Oy/AIR CYL-
INDER (PIN-INDEX) GAS SUPPLY ASSEMBLY, PARTS LIST ADDS THE FOLLOWING:
Gas Supply, Oo/Air Pipeline (NiST) with Cylinder (Pin-Index) Low Pressure - Part Number 81 501 95.
Add to Item 23 regulator, 45 psi W/press relief
~ Part number 81 501 77.