SETTING OF DELAY OF DOOR ALARM
The door check indicator is light when the door is opened, and the alarm buzzer sounds with some delay
to notice the door opening.
The delay time (between lighting of the door check indicator and activation of the alarm buzzer) can be
changed.
Set an appropriate delay time according to the condition of use to prevent the rise of chamber
temperature resulting from inadequate door close.
■
Setting range of delay time:
1 and 15 minutes
■
Initial setting (factory setting):
2 minutes
Table 5
Procedure for setting of delay of door alarm (Change the delay time to 3 min. from 2
min )___
Description of operation
Key operated
Indication after operation
1
The current chamber temperature is
displayed.
4
2
Press up arrow key for 5 seconds.
1
F00 is displayed and the first digit
blinks.
F00
'IV
3
Press up arrow key 4 times.
1
The display is changed to F04 from
F00.
F0H
4
Press set key.
SET
The current setting (002) is displayed
and the first digit blinks.
0133
5
Press up arrow key and scroll the
figure to 3.
i
The display is changed to 003 from
002.
\l/
003
✓
iv
6
Press set key.
SET
Delay time is memorized and the
current chamber temperature is
displayed.
4
Note:
The setting mode returns to the temperature display mode automatically when 90 seconds has passed
without any key operation (auto-return function).
In this case, the setting is not accepted.
—
56
—
18