TruSense Water Chemistry Analyzer | Owner's Manual
®
4.
Turn off pump and filtration system at the breaker. Automation system
needs to be in auto NOT service mode.
5.
Close the valves if needed.
6.
Disconnect the Blue pH sensor BNC connector from the BLUE labeled
left side connector of the electronic circuit board
7.
Loosen the left compression fitting and lift pH sensor carefully from
plastic housing, DO NOT touch the glass tube.
8.
Connect the Blue pH sensor BNC connector to the BLUE labeled left
side connector of the electronic circuit board.
9.
Submerge the sensor into the container with measured water sample.
Start calibration using the iAquaLink app or on the OneTouch
controller (AquaLink
SETUP > TRUSENSE, see Figure 10.
Figure 10.
Submerge sensor in water sample
10. Remove the pH sensor from storage tube and rinse the tip of the sensor
with clean water.
11. Once calibration is complete, carefully insert pH sensor back into
Jandy TruSense and hand tighten the compression fitting.
RS only) go to: MENU / HELP > SYSTEM
®
Calibrate pH
Page 17