5.
Loosen the blue left compression fitting and carefully lift pH sensor
from the plastic housing, DO NOT touch the glass tube,
see Figure 6.
Figure 6.
Remove pH sensor
6.
Disconnect the Blue pH sensor BNC connector from the BLUE labeled
left side connector of the electronic circuit board.
7.
Rinse the tip of the pH sensor with clean water.
8.
Shake pH sensor to remove residual water. DO NOT put the sensor on
the ground or touch the glass tube.
9.
Connect the Blue pH sensor BNC connector to the BLUE labeled left
side connector of the electronic circuit board.
NOTE: The pH sensor wire should remain connected during calibration.
10. Submerge the sensor into the pH 7 buffer solution inside the storage
tube. Start calibration in the iAquaLink app or on the OneTouch
controller (AquaLink
SETUP > pH > 2 POINT, see Figure 7.
Jandy
TruSense Water Chemistry Analyzer | Owner's Manual
®
RS only) go to: MENU / HELP > SYSTEM
®