profometer PM8500 Quick Start Manual page 11

Hide thumbs Also See for PM8500:

Advertisement

PM8500 Wheel Electrodes
Filling the wheel electrodes with Copper Sulphate
A
2
The wooden plug (A) is not intended to be removed. Immerse the wheel in water for an hour to allow the water to soak into the plug.
a
Only the wheels should be immersed, therefore, please remove them from the legs before doing it. 
The felt hoops (B) and the felt hoop connector must be saturated before measuring. In normal use (with water and cupper solution),
b
the thin felt is on the wooden spigot side (A).
Prepare the saturated solution by mixing 40 units by weight of copper sulphate with 100 units by weight of distilled water. Fill in the
c
cupper tank removing the plug (C). Add an additional teaspoon of copper sulphate crystals into the electrode.
The electrode should be filled as completely as possible with a minimum of air in the compartment. 
d
Fill in the water container with water. To do so you need to remove the water plug (D).
The water should be dripping out slowly from the water tank throughout the 1mm nozzle (E). The bike valve on the opposite side
e
should provide a high pressure; if water is dripping out too fast, please apply some pressure blowing the bike valve with your mouth.
E
B
D
C