Summary of Contents for PCB Piezotronics M1381-001-01A
Page 1
Model M1381-001-01A rod end load cell Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free: 800-828-8840 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com Web: www.pcb.com...
Page 2
Piezotronics, user servicing or repair is critical test. not recommended and, if attempted, may void the factory warranty. Routine PCB Piezotronics maintains an ISO- maintenance, such as the cleaning of 9001 certified metrology laboratory and electrical connectors, housings, and offers calibration services, which are...
Page 3
50% of the general contact numbers are: replacement cost returned item(s). PCB will provide a price PCB Piezotronics, Inc. quotation replacement 3425 Walden Ave. recommendation for any item whose Depew, NY14043 USA repair costs would exceed 50% of...
ROD END LOAD CELL OPERATION MANUAL 3.1 Dimensions 1.0 INTRODUCTION Rod end load cells manufactured by PCB Load & Torque, Inc. are designed for integration into tension/compression measurement applications such as process automation, quality assurance, and production monitoring. Standard 3/4”-16 and 1”-14 male/female threads facilitate ease of installation.
ROD END LOAD CELL OPERATION MANUAL 5.3 Cable & Grounding Considerations 5.0 ELECTRICAL INSTALLATION Proper grounding and shielding is required to prevent Table 4 - Electrical Connections electrical noise in strain gage load cell measuring systems. The cable must be shielded twisted pairs with a drain wire. Load Cell Receptacle: PT02E-10-6P Mating Connector:...
Page 9
ROD END LOAD CELL OPERATION MANUAL 6.1.1 Measured Output 7.0 SHUNT CALIBRATION DESCRIPTION The applied load starting at zero is measured in five Shunt calibration is used to simulate a known tension or increments to full scale. Output (mV/V) is measured at each compression load on a load cell.
ROD END LOAD CELL OPERATION MANUAL 8.0 MAINTENANCE 10. Keep a record of your observations, correct problems, or contact PCB factory for assistance. Routine maintenance of the rod end load cell should include cleaning the electrical connectors, housings, and mounting surfaces with solutions and techniques that will not harm the 10.0 CALIBRATION / REPAIR SERVICES physical material of construction.
Page 11
Model Number Revision: B ROD END LOAD CELL M1381-001-01A ECN #: 43827 Performance ENGLISH OPTIONAL VERSIONS Measurement Range 5000 lbf 22 kN Optional versions have identical specifications and accessories as listed for the standard Sensitivity(± 10 %) 2 mV/V 2 mV/V model except where noted below.
Page 12
1.50 [38.1] Ø 1.27 [32.3] "A" THREAD 1.00 [25.4] BOTH ENDS 1.30 [32.9] FLATS 1.03 [26.2] 1.11 [28.2] FLATS M1381-001-01A M14x1.5 - 6H 4.00 [101.6] M1381-002-01A M16x1.5 - 6H MODEL NUMBER "A" THREAD UNLESS OTHERWISE SPECIFIED TOLERANCES ARE: DRAWN CHECKED ENGINEER WIRING CODE FOR MODELS WITH "TEDS"...
Need help?
Do you have a question about the M1381-001-01A and is the answer not in the manual?
Questions and answers