PCB Piezotronics 480M122 Installation And Operating Manual

Battery-powered signal conditioner

Advertisement

Quick Links

Model 480M122
Battery-Powered Signal Conditioner
Installation and Operating Manual
For assistance with the operation of this product,
contact PCB Piezotronics, Inc.
Toll-free: 800-828-8840
24-hour SensorLine: 716-684-0001
Fax: 716-684-0987
E-mail: info@pcb.com
Web: www.pcb.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 480M122 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for PCB Piezotronics 480M122

  • Page 1 Model 480M122 Battery-Powered Signal Conditioner Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free: 800-828-8840 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com Web: www.pcb.com...
  • Page 2 Piezotronics, user servicing or repair is critical test. not recommended and, if attempted, may void the factory warranty. Routine PCB Piezotronics maintains an ISO- maintenance, such as the cleaning of 9001 certified metrology laboratory and electrical connectors, housings, and offers calibration services, which are...
  • Page 3 50% of the general contact numbers are: replacement cost returned item(s). PCB will provide a price PCB Piezotronics, Inc. quotation replacement 3425 Walden Ave. recommendation for any item whose Depew, NY14043 USA repair costs would exceed 50% of...
  • Page 4 PCB工业监视和测量设备 - 中国RoHS2公布表 PCB Industrial Monitoring and Measuring Equipment - China RoHS 2 Disclosure Table 有害物质 汞 镉 六价铬 (Cr(VI)) 多溴联苯 (PBB) 多溴二苯醚 (PBDE) 部件名称 铅 (Pb) (Hg) (Cd) 住房 PCB板 电气连接器 压电晶体 环氧 铁氟龙 电子 厚膜基板 电线 电缆 塑料 焊接...
  • Page 5 Component Name Hazardous Substances Lead Mercury Cadmium Chromium VI Polybrominated Polybrominated (Pb) (Hg) (Cd) Compounds Biphenyls Diphenyl (Cr(VI)) (PBB) Ethers (PBDE) Housing PCB Board Electrical Connectors Piezoelectric Crystals Epoxy Teflon Electronics Thick Film Substrate Wires Cables Plastic Solder Copper Alloy/Brass This table is prepared in accordance with the provisions of SJ/T 11364.
  • Page 6 Model 482C02 / 480E09 ® Battery-Powered ICP Signal Conditioners Operating Manual with Enclosed Warranty Information 3425 Walden Avenue, Depew, New York 14043-2495 Phone (716) 684-0003 Fax (716) 684-3823 Toll Free Line 1-800-828-8840 MANUAL NUMBER: 19175 MANUAL REVISION: A ECO: 50452...
  • Page 7: Table Of Contents

    Table of Contents Introduction ........................page 3 Description ........................page 4 Operation .......................... page 5 Performance Limits ......................page 8 Battery Considerations ..................... page 8 Optional Power Connections .................... page 9 Warranty ......................... page 10 Calibration & Service ...................... page 10 Return Procedure ......................
  • Page 8: Introduction

    Introduction ® Model 480C02 & model 480E09 are rugged, portable, power sources for ICP sensors. These models are nearly identical except model 480C02 is unity gain 1:1 only with no adjustment, while model 480E09 has the additional feature of 3-position gain adjustment (1/10/100). They are powered internally by three replaceable batteries and ®...
  • Page 9: Description

    Red (fault, electrical short) Yellow (fault, low battery) EXT PWR – use with 488A03 AC Power Source Gain Select – Rotary Switch Green (connected & no fault) Figure 2: Model 480E09 Signal Conditioner Description Models 480C02 / 480E09 contain three, PP3 size, 9 volt batteries connected in series to provide DC power to internal electronics which provide regulation for approximately 2.6 mA of constant current power.
  • Page 10: Operation

    Operation Internal electronics include a capacitor and resistor that decouples the sensor output signal from the DC power. With no sensor connected, move power switch to “ON” position. The front panel voltmeter measures the battery ® voltage with full scale set at 27 volts (+27 volts for new batteries), see Figure 1. When an ICP sensor is connector to the input “XDCR”...
  • Page 11 If the sensor amplifier and/or cable are damaged to short open, the meter will indicate in the full scale (yellow) area. Should the sensor amplifier and/or cable be damaged to short closed, the meter will indicate zero volts (red area). Immediately after connecting readout instrument, (oscilloscope, meter, recorder, etc.) to the output jack, the 47 F coupling capacitor will begin charging through the internal resistor and input resistance of the readout instrument.
  • Page 12 DC Coupling for Low-Frequency Response With the 480C02 / 480E09 connected as shown in Figure 5, the low frequency response of the coupling circuit is determined by the relationship in Equation 2. This requires use of a BNC Tee. Figure 6 below provides a comparison of AC vs DC coupling the same signal.
  • Page 13: Performance Limits

    Performance Limits Output Voltage Limits Certain ICP sensors are capable of 10-volt output voltage swing. Models 480C02 / 480E09 with 27 volt supply will allow the positive-going side of the signal to go to +14 volts. The negative signal side is capable of -8 volts assuming a 10 volt turn-on for the sensor.
  • Page 14: Optional Power Connections

    Battery Life & Replacement Typical battery life is 100 hours at gain of 1 or 40 hours at gain of 10 or 100 when using high quality alkaline batteries, assuming the unit is turned off when not in use. When a BATT Test fail occurs, find and carefully remove the Philips head screw on the back of model 480C02.
  • Page 15: Warranty

    Warranty PCB instrumentation is warranted against defective material and workmanship for 1 year unless otherwise expressly specified. Damage to instruments caused by incorrect power or misapplication, is not covered by warranty. If there are any questions regarding power, intended application, or general usage, please consult with your local sales contact or distributor.
  • Page 16: Total Customer Satisfaction

    Total Customer Satisfaction PCB, a division of PCB Piezotronics, guarantees Total Customer Satisfaction. If, at any time, for any reason, you are not completely satisfied with any PCB product, PCB will repair, replace, or exchange it at no charge. You may also choose, within the warranty period, to have your purchase price refunded.
  • Page 17 Model Number Revision: F BATTERY-POWERED SIGNAL CONDITIONER 480M122 ECN #: 45339 Performance ENGLISH OPTIONAL VERSIONS Channels Optional versions have identical specifications and accessories as listed for the standard Frequency Range(-5 %)(x1, x10 Gain) 0.15 to 100,000 Hz 0.15 to 100,000 Hz model except where noted below.

This manual is also suitable for:

480c02

Table of Contents