Summary of Contents for PCB Piezotronics ICP 109B01
Page 1
High freq. ICP® pressure sensor, 80k psi, 0.07 mV/psi, accel comp., integral thd (ceramic coated diaphragm) Installation and Operating Manual For assistance with the operation of this product, contact the PCB Piezotronics, Inc. Toll-free: 716-684-0001 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com...
Assistance is needed to safely operate equipment PCB Piezotronics is an ISO-9001 certified company whose Damage is visible or suspected calibration services are accredited by A2LA to ISO/IEC Equipment fails or malfunctions 17025, with full traceability to SI through N.I.S.T.
Page 3
CAUTION Refers to hazards that could damage the instrument. NOTE Indicates tips, recommendations and important information. The notes simplify processes and contain additional information on particular operating steps. The following symbols may be found on the equipment described in this manual: This symbol on the unit indicates that high voltage may be present.
OPERATION MANUAL FOR HIGH PRESSURE SENSORS ® Series 108 & 109 INTRODUCTION low-impedance, high-level output signal. See the Technical Information section of our website for The 108 & 109 Series are acceleration-compensated, ® more complete coverage of ICP instruments. ® high pressure, ICP (Integrated Circuit Piezoelectric) sensors primarily developed to measure ballistics...
OPERATION MANUAL FOR HIGH PRESSURE SENSORS ® Series 108 & 109 The limitation in this type of installation lies in the MODEL 040A frequency-limiting effects of the passage due to its REAMER length. MODEL 041A The passage behaves like an underdamped second PILOT BUSHING order system;...
OPERATION MANUAL FOR HIGH PRESSURE SENSORS ® Series 108 & 109 FLASH TEMPERATURE PASSAGE PASSAGE APPROX. EFFECTS LENGTH (in) RESONANCE FASTEST PULSE If more protection from flash thermal effects is (kHz) RISE TIME (sec) required with the recessed mount, the passage can be .050 filled with silicone grease (DC-4 or equivalent).
OPERATION MANUAL FOR HIGH PRESSURE SENSORS ® Series 108 & 109 CIRCUIT CONNECTIONS 002 SERIES The figures below show two typical circuit ® connections normally used to connect the 108 & 109 POWER SUPPLY Series to power units. The first figure is the connection scheme used for READOUT most applications.
MAINTENANCE The small size and sealed construction of the 108 & 109 Series precludes field maintenance and repair. Contact the factory for further assistance. ® is a registered trademark of PCB Piezotronics Drawing Number: 21113 Revision: B...
Page 12
PCB Piezotronics Inc. claims proprietary rights in REVISIONS the information disclosed hereon. Neither it nor any reproduction thereof will be disclosed to others DESCRIPTION without the written consent of PCB Piezotronics Inc. REVISED DIMENSION .244-.250, ADDED CERAMIC 53103 ELECTRICAL CONNECTOR COATING NOTE...
Need help?
Do you have a question about the ICP 109B01 and is the answer not in the manual?