Summary of Contents for PCB Piezotronics ICP 108B03
Page 1
Model 108B03 ICP® Pressure Sensor Installation and Operating Manual For assistance with the operation of this product, contact PCB Piezotronics, Inc. Toll-free: 800-828-8840 24-hour SensorLine: 716-684-0001 Fax: 716-684-0987 E-mail: info@pcb.com Web: www.pcb.com...
Page 2
Piezotronics, user servicing or repair is critical test. not recommended and, if attempted, may void the factory warranty. Routine PCB Piezotronics maintains an ISO- maintenance, such as the cleaning of 9001 certified metrology laboratory and electrical connectors, housings, and offers calibration services, which are...
Page 3
50% of the general contact numbers are: replacement cost returned item(s). PCB will provide a price PCB Piezotronics, Inc. quotation replacement 3425 Walden Ave. recommendation for any item whose Depew, NY14043 USA repair costs would exceed 50% of...
OPERATION MANUAL FOR ® HIGH PRESSURE SENSORS Series 108 & 109 The microelectronic amplifier converts the high- INTRODUCTION impedance voltage from the quartz package into a The 108 & 109 Series are acceleration-compensated, low-impedance, high-level output signal. See the ® high pressure, ICP (Integrated Circuit Piezoelectric) Technical Information section of our website for...
OPERATION MANUAL FOR ® HIGH PRESSURE SENSORS Series 108 & 109 The limitation in this type of installation lies in the MODEL 040A frequency-limiting effects of the passage due to its REAMER length. MODEL 041A The passage behaves like an underdamped second PILOT BUSHING order system;...
OPERATION MANUAL FOR ® HIGH PRESSURE SENSORS Series 108 & 109 The natural frequency and approximate fastest FLASH TEMPERATURE pressure step rise time for various length passages is EFFECTS shown in the following chart. (Medium, air at 25C). The ceramic coating on the diaphragm of these sensors should render the flash thermal effect PASSAGE PASSAGE...
OPERATION MANUAL FOR ® HIGH PRESSURE SENSORS Series 108 & 109 CIRCUIT CONNECTIONS 002 SERIES The figures below show two typical circuit ® connections normally used to connect the 108 & 109 POWER SUPPLY Series to power units. The first figure is the connection scheme used for READOUT most applications.
MAINTENANCE The small size and sealed construction of the 108 & 109 Series precludes field maintenance and repair. Contact the factory for further assistance. ® is a registered trademark of PCB Piezotronics Drawing Number: 21113 Revision: A...