This type of warning is used to indicate an imminent risk that can cause serious injury or death if instructions are not followed. Alcatel Vacuum Technology France - User’s Manual ACC 2009 1/15...
Any modification could cause damage or personnel injury. Skilled personnel All work described in this document should be carried out only by qualified persons who have suitable technical training and the necessary experience. Alcatel Vacuum Technology France - User’s Manual ACC 2009 2/15...
Operating position Any orientation Internal volume 18.3 cm Receivable maximum pressure 2 bars Materials exposed to vacuum Molybdenum, SUS304, SUS430 Tungsten, , Fe / Ni+Ni-planting, Glass, FPM Filament material Tungsten (W) Alcatel Vacuum Technology France - User’s Manual ACC 2009 4/15...
Page 6
Install a circuit breaker on the power source. Disregarding these instructions can cause damage and/or injury. (1) ”D-Sub” 9P (female) connector is available to allow the gauge to be connected directly to customer equipment. ( 3 p. 3). Alcatel Vacuum Technology France - User’s Manual ACC 2009 5/15...
This gauge is designed to operate in any orientation. Install the gauge horizontally for higher accuracy. Install this gauge away from drastic temperature changes (away from any heat sources or fans) for higher accuracy. Alcatel Vacuum Technology France - User’s Manual ACC 2009 6/15...
ACC 2009 - 0/10 V + SP use to enter set point data ( 6 p. 10) ACC 2009 - RS485 + SP use to enter network address (refer to RS485 Communication Manual) Alcatel Vacuum Technology France - User’s Manual ACC 2009 7/15...
Page 9
4,1 0,8 0,5 0,4 RS485 Signal output RS485 communication is available on the connector Pins 1(+) and 8(–) . (ACC 2009 - RS485 + SP) Refer to the RS485 Communication Manual for details. Alcatel Vacuum Technology France - User’s Manual ACC 2009 8/15...
Page 10
Follow this example to set up the set points (refer also to table next page): set point 1: 5x10 = 500 mbar –1 set point 2: 8x10 = 0.8 mbar Alcatel Vacuum Technology France - User’s Manual ACC 2009 9/15...
Page 11
– End of process – Note: For the ACC2009 - RS485 + SP models, the set points are modified only via the RS485 interface (refer to the RS485 Communication Manual). Alcatel Vacuum Technology France - User’s Manual ACC 2009 10/15...
- Place a protective cap on the fitting to prevent foreign substances from entering the gauge. - If the gauge must be stored, we advise to keep it away from heat, moisture an dust ( 4 p. 4). Alcatel Vacuum Technology France - User’s Manual ACC 2009 11/15...
- If the measured value is out of this range, replace the sensor. 57 Ω Maintenance of the The items required for maintenance of the sensor are available sensor in a kit. Alcatel Vacuum Technology France - User’s Manual ACC 2009 12/15...
Page 14
V < 8,582 V surements Replace sensor head green sensor contaminated V < 2,199 V Err06 Electronic error Replace electronics V < 0,5 V ErrL0 Electronic error Replace electronics 0,2 V Alcatel Vacuum Technology France - User’s Manual ACC 2009 13/15...
The manufacturer’s obligation to recover such equipment applies only to complete items of equipment that have been neither modified nor retrofitted and have used only spare parts from Alcatel Vacuum Technology sold by Alcatel Vacuum Technology and including all their assemblies and sub-assemblies. This obligation does not include the cost of transporting the product to a reprocessing centre, nor the cost of the service which will be invoiced on to the customer.