Alcatel Vacuum Technology adixen ACC 2009 User Manual

Alcatel Vacuum Technology adixen ACC 2009 User Manual

Cold cathode/pirani gauge

Advertisement

Cold Cathode/Pirani Gauge
ACC 2009
User's Manual
Alcatel Vacuum Technology France - 98, avenue de Brogny - BP 2069 - 74009 Annecy cedex - FRANCE
Tel. (33) 4 50 65 77 77 - Fax. (33) 4 50 65 77 89
Web site: www.adixen.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the adixen ACC 2009 and is the answer not in the manual?

Questions and answers

Summary of Contents for Alcatel Vacuum Technology adixen ACC 2009

  • Page 1 ACC 2009 User’s Manual Alcatel Vacuum Technology France - 98, avenue de Brogny - BP 2069 - 74009 Annecy cedex - FRANCE Tel. (33) 4 50 65 77 77 - Fax. (33) 4 50 65 77 89 Web site: www.adixen.com...
  • Page 2: Table Of Contents

    This type of warning is used to indicate an imminent risk that can cause serious injury or death if instructions are not followed. Alcatel Vacuum Technology France - User’s Manual ACC 2009 1/15...
  • Page 3: Safety

    Any modification could cause damage or personnel injury. Skilled personnel All work described in this document should be carried out only by qualified persons who have suitable technical training and the necessary experience. Alcatel Vacuum Technology France - User’s Manual ACC 2009 2/15...
  • Page 4: Responsibility And Warranty

    - Accessories Description Gauge cable length (5 m) 112752 Gauge cable length (10 m) 112753 Gauge cable length (20 m) 112754 “D-Sub“ 9P (female) 114848 connector (for custom equipment connections) Alcatel Vacuum Technology France - User’s Manual ACC 2009 3/15...
  • Page 5: Technical Data

    Operating position Any orientation Internal volume 18.3 cm Receivable maximum pressure 2 bars Materials exposed to vacuum Molybdenum, SUS304, SUS430 Tungsten, , Fe / Ni+Ni-planting, Glass, FPM Filament material Tungsten (W) Alcatel Vacuum Technology France - User’s Manual ACC 2009 4/15...
  • Page 6 Install a circuit breaker on the power source. Disregarding these instructions can cause damage and/or injury. (1) ”D-Sub” 9P (female) connector is available to allow the gauge to be connected directly to customer equipment. ( 3 p. 3). Alcatel Vacuum Technology France - User’s Manual ACC 2009 5/15...
  • Page 7: Gauge Installation

    This gauge is designed to operate in any orientation. Install the gauge horizontally for higher accuracy. Install this gauge away from drastic temperature changes (away from any heat sources or fans) for higher accuracy. Alcatel Vacuum Technology France - User’s Manual ACC 2009 6/15...
  • Page 8: Operation

    ACC 2009 - 0/10 V + SP use to enter set point data ( 6 p. 10) ACC 2009 - RS485 + SP use to enter network address (refer to RS485 Communication Manual) Alcatel Vacuum Technology France - User’s Manual ACC 2009 7/15...
  • Page 9 4,1 0,8 0,5 0,4 RS485 Signal output RS485 communication is available on the connector Pins 1(+) and 8(–) . (ACC 2009 - RS485 + SP) Refer to the RS485 Communication Manual for details. Alcatel Vacuum Technology France - User’s Manual ACC 2009 8/15...
  • Page 10 Follow this example to set up the set points (refer also to table next page): set point 1: 5x10 = 500 mbar –1 set point 2: 8x10 = 0.8 mbar Alcatel Vacuum Technology France - User’s Manual ACC 2009 9/15...
  • Page 11 – End of process – Note: For the ACC2009 - RS485 + SP models, the set points are modified only via the RS485 interface (refer to the RS485 Communication Manual). Alcatel Vacuum Technology France - User’s Manual ACC 2009 10/15...
  • Page 12: Gauge Removal

    - Place a protective cap on the fitting to prevent foreign substances from entering the gauge. - If the gauge must be stored, we advise to keep it away from heat, moisture an dust ( 4 p. 4). Alcatel Vacuum Technology France - User’s Manual ACC 2009 11/15...
  • Page 13: Maintenance

    - If the measured value is out of this range, replace the sensor. 57 Ω Maintenance of the The items required for maintenance of the sensor are available sensor in a kit. Alcatel Vacuum Technology France - User’s Manual ACC 2009 12/15...
  • Page 14 V < 8,582 V surements Replace sensor head green sensor contaminated V < 2,199 V Err06 Electronic error Replace electronics V < 0,5 V ErrL0 Electronic error Replace electronics 0,2 V Alcatel Vacuum Technology France - User’s Manual ACC 2009 13/15...
  • Page 15: Returning The Products

    The manufacturer’s obligation to recover such equipment applies only to complete items of equipment that have been neither modified nor retrofitted and have used only spare parts from Alcatel Vacuum Technology sold by Alcatel Vacuum Technology and including all their assemblies and sub-assemblies. This obligation does not include the cost of transporting the product to a reprocessing centre, nor the cost of the service which will be invoiced on to the customer.
  • Page 16: Declaration Of Conformity

    - Declaration of Conformity Alcatel Vacuum Technology France - User’s Manual ACC 2009 15/15...

Table of Contents