Download Print this page
Hanna Instruments HI 762 Manual

Hanna Instruments HI 762 Manual

Free chlorine ultra low range

Advertisement

Quick Links

HI 762
Free Chlorine Ultra Low Range
w w w . h a n n a i n s t . c o m
Dear Customer,
Thank you for choosing a Hanna Instruments Product.
Please read this instruction manual carefully before using the instrument. If you need
additional technical information, do not hesitate to e-mail us at tech@hannainst.com
or view our worldwide contact list at www.hannainst.com.
Preliminary examination:
Please examine this product carefully. Make sure that the instrument is not damaged. If
any damage occurred during shipment, please notify your Dealer.
Each HI 762 meter is supplied complete with:
• Two Sample Cuvettes and Caps
• Reagent set for 6 tests
• 1 x 1.5V AAA Battery
• Instruction Manual and Quick Reference
For more details about spare parts and accessories
see "Accessories".
Technical specifications:
Range
0 to 500 ppb
1 ppb
Resolution
Accuracy
±20 ppb ±4% of reading @ 25 °C / 77 °F
Light Source
Light Emitting Diode @ 525 nm
Light Detector
Silicon Photocell
Method
Adaptation of the USEPA method 330.5. The reaction
between free chlorine and DPD reagent causes a pink
tint in the sample.
Environment
0 to 50 °C (32 to 122 °F);
max 95% RH non-condensing
Battery Type
1 x 1.5V AAA
Auto-Shut off
After 10 minutes of non-use
Dimensions
86.0 x 61.0 x 37.5 mm (3.4 x 2.4 x 1.5")
Weight
64 g (2.3 oz.)
Functional description:
1. Dust cover.
2. Cuvette with cap.
3. Cuvette holder.
4. Liquid Crystal Display.
5. Button.
Errors and warnings:
Light High: There is too much light to perform a measurement.
Please check the preparation of the zero cuvette.
Light Low: There is not enough light to perform a measurement.
Please check the preparation of the zero cuvette.
Inverted Cuvettes: The sample and the zero cuvette are inverted.
Under Range: A blinking "0" indicates that the sample absorbs less
light than the zero reference. Check the procedure and make sure you
use the same cuvette for reference (zero) and measurement.
Over Range: A flashing value of the maximum concentration indicates
the reading is over range. Dilute the sample and re-run the test.
Battery Low: The battery must be replaced soon.
Dead Battery: This indicates that the battery is dead and must be
replaced. Once this indication is displayed, normal operation of the
instrument will be interrupted. Change the battery and restart the
meter.

Advertisement

loading

Summary of Contents for Hanna Instruments HI 762

  • Page 1 Please examine this product carefully. Make sure that the instrument is not damaged. If 4. Liquid Crystal Display. any damage occurred during shipment, please notify your Dealer. Each HI 762 meter is supplied complete with: 5. Button. • Two Sample Cuvettes and Caps •...
  • Page 2 • Turn the instrument off by holding the button until the meter shuts off. unscrew the cap. Add the content of one Hanna Instruments reserves the right to modify the design, construction or • Turn the instrument upside down and remove the battery cover with a screwdriver.