HP 41951A Operation Note page 33

Hide thumbs Also See for 41951A:
Table of Contents

Advertisement

SECTION 6
SERVICE
6-1. INTRODUCTION
This section provides the information required to service the 41951A Impedance Test Kit.
The Schematic Diagram, and the Replaceable Parts List are included.
6-2. SCHEMATICS AND REPLACEABLE PARTS
The 41951A consists of the assemblies
listed in Table 1-2.
Any assembly except for the
impedance Test Adapter, can be replaced by the assembly level, when it is damaged.
The
part numbers are shown in Figure 1-1, and Table 1-2.
The
impedance
Test Adapter
includes
some
replaceable
parts.
Figure 6-1
shows
the
schematic diagram of the Impedance Test Adapter.
The replaceable parts mounted on the
board in the Impedance Test Adapter, is listed in Table 6-1.
The part's locations are shown
in Figure 6-2.
NOTE
Parts that are not listed in Table 6-1 cannot be replaced, and the board assembly
cannot be replaced.
So if any of parts which are not listed are damaged, the
impedance Test Adapter must be replaced at the assembly level.
Table 6-2 shows an exploded view of the APC-7 connector mounted on the Impedance Test
Adapter's top cover, and lists the APC-7 connecter's parts.
Table 6-3 lists the Impedance
Test Adapter's replaceable parts except for those on the board, and shows their location.

Advertisement

Table of Contents
loading

Table of Contents