i-Flow 7000 Series N2 Generators
User Manual
Copyright © Peak Scientific Instruments Ltd 2023 — i-Flow 7000 Series User Manual Rev 1 RSID 4707 EN - Alt —13/01/2023

Advertisement

Table of Contents
loading

Summary of Contents for Peak Gas Generation i-Flow 7000 Series

  • Page 1 7000 Series N2 Generators User Manual Copyright © Peak Scientific Instruments Ltd 2023 — i-Flow 7000 Series User Manual Rev 1 RSID 4707 EN - Alt —13/01/2023...
  • Page 2 Register Product to Operate To begin operation you will need to register your generator. You can do so by visiting www.peakscientific.com/activate or by downloading the Peak Genius App. Registering will activate your 2 year warranty* - covering every single component in your generator. ...
  • Page 3: Table Of Contents

    Contents Change History Introduction Technical Description Basic Concept i-Flow 7000 Series Model Number Breakdown Warranties and Liabilities Warranty & Liability Coverage Safety Notices Symbols Safety Notice to Users EU Declaration of Conformity CSA Compliance Statement WEEE Compliance Statement Technical Specification...
  • Page 4: Change History

    Change History Comment Name Date How to use this Manual This manual is intended for end users and has been written as a reference document where you can skip to the relevant information. Users can refer to the contents page to find the relevant information. Please review each of the following sections carefully.
  • Page 5: Introduction

    Peak Gas Generation or the Peak Partner from which you purchased your Generator. Peak Gas Generation is a Trading Name of Peak Scientific Instruments Ltd. All Peak Gas Generation products are manufactured by Peak Scientific Instruments Ltd.
  • Page 6: Technical Description

    Technical Description Basic Concept i-Flow N2 Mini & N2 Prime are a range of modular nitrogen gas generators, that operate based on the latest Pressure Swing Adsorption (PSA) technology, utilising a Carbon Molecular Sieve (CMS) to extract nitrogen from air, at the specified flow, purity & pressure.
  • Page 7: I-Flow 7000 Series Model Number Breakdown

    CMS columns can be added. (Note: Mini models with 3-4 & Prime models with 5-10 banks require additional vent valves & silencers fitted to the rear of CMS column assembly). i-Flow 7000 Series Model Number Breakdown Series Name Customer Option No.
  • Page 8: Warranties And Liabilities

    Warranties and Liabilities Warranty & Liability Coverage Peak warrants that, subject to the provisions in this statement, purchased Peak generators, whether purchased directly from Peak or indirectly via an approved, certified and trained distributor or partner (referred to hereafter as a “Peak Partner”) will comply in all material respects with any specifications referred to in your customer order confirmation and, subject to installation and operational guidelines being followed as described in applicable product manuals, shall be free from any defects in quality of materials or workmanship for a period of one year from the date of installation, provided this takes...
  • Page 9 8. Peak shall not be liable in respect of any claim made for costs, damages, losses or expenses (whether consequential, direct, indirect or otherwise) or in any respect howsoever arising including, but not limited to, liability from accident or negligence (other than pursuant to clause 5) that may be suffered by you or any third party.
  • Page 10: Safety Notices

    Safety Notice to Users These instructions must be read thoroughly and understood before installation and operation of your Peak i-Flow 7000 Series Gas Generator. Use of the Generator in a manner not specified by Peak Gas Generation MAY impair the built in SAFETY features of the equipment.
  • Page 11: Eu Declaration Of Conformity

    EU Declaration of Conformity Peak Scientific Instruments Ltd. Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE Hereby declare that, this declaration of conformity is issued under the sole responsibility of the manufacturer. Equipment Type: Nitrogen Generator Model Designator: i-Flow 7000 & 9000 Series PED Module B Certificate No : CE 744518 PED Module D Certificate No : CE 608103 Pressure assembly consisting of the following key parts :...
  • Page 12: Csa Compliance Statement

    CSA Compliance Statement CSA Group (Canadian Standards Authority) is a Nationally Recognised Testing Laboratory (NRTL), headquartered in Toronto Canada. They are authorised to evaluate product to both their own and Underwriters Laboratory (UL) standards and certify the product to be in compliance to the relevant standards. Peak products are certified to the current in force revision of the following standards in order to cover both Canadian and United States requirements for “Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory use,...
  • Page 13: Weee Compliance Statement

    WEEE Compliance Statement The Waste Electrical and Electronic Equipment (WEEE) Regulations SI 2013 No 3113 and or the Waste Electrical and Electronic Equipment (WEEE) Directive 2012/19/EU apply to all electrical and electronic equipment placed on the market in the UK and EU covered by the scope of regulations which can be found in the Government Guidance Notes (PDF) produced by the Department for Business Innovation and skills for the UK and here for Europe.
  • Page 14: Technical Specification

    Depending on your specific application, the pressure and flow required from the compressed air supply to the generator will vary. For your specific requirements please refer to your quotation documents, or contact Peak Gas Generation for further information. However all installations must meet the following conditions: Minimum Air Quality IS0 8573-1:2010 class 2.4.1...
  • Page 15 General i-Flow Mini 1 Bank 2 Banks 3 Banks 4 Banks Width mm(in) 418 (16.5) Height mm(in) 853 (33.6) Depth mm(in) 830 (32.7) 992 (39.1) 1154 (45.4) 1316 (51.8) Weight kg(lbs) 96 ( 211) 132 (291) 178 (392) 214 (471) Shipping weight kg(lbs) 115 (253) 153 (337)
  • Page 16: Unpacking

    Check ‘SHOCKWATCH’ and ‘TIP-N-TELL’ labels for signs of rough handling prior to unpacking. Any damage should be reported immediately to the carrier and Peak Gas Generation or the Peak Partner from where the unit was purchased. Page 16...
  • Page 17: Installation

    Installation It is the user/installer’s responsibility to ensure that the generator is located and protected against any external influences such as vibration, shock, wind, snow loading, earthquake or fire. The installation should conform to all local regulations and should be leak tight and completed by technically competent personnel.
  • Page 18: Generator Overview

    Generator Overview i-Flow Mini General Dimensions 418 mm / 16.4” 830 mm / 32.7” 1316 mm / 51.8” Note: Each additional bank assembly = 162mm / 6.4” The Generator must always be placed on a level surface. Failure to do so will affect the stability of the Generator.
  • Page 19: Generator Overview

    Generator Overview i-Flow Prime General Dimensions 418 mm / 16.4” 730 mm / 28.7” 2188 mm / 86.1” Note: Each additional bank assembly = 162mm / 6.4” The Generator must always be placed on a level surface. Failure to do so will affect the stability of the Generator.
  • Page 20: Inlet / Outlet Connections

    Inlet / Outlet Connections The i-Flow 7000 Series generators can be piped (air inlet and gas outlet) from either side to suit site/ location. The unit will be supplied with two blanking plugs which can be installed in either the left or right hand ports to allow the opposite side to be used.
  • Page 21: Recommended Piping Layout

    Recommended Piping Layout To allow for proper operation and commissioning of the generator it is important to include isolation and vent valves at the locations displayed below, the i-Flow 7000 Series generators can be piped (air inlet and gas outlet) from either side to suit site/ location.
  • Page 22: Compressed Air Quality

    WARNING i-Flow 7000 series nitrogen generators will typically be the last stage in a complete air compression and filtration system, the inlet air quality must meet a minimum of IS0 8573-1:2010 class 2.4.1 or higher. Class 2.4.1 is defined as.
  • Page 23: Electrical Connection

    Electrical Connection Connect the Generator to a single-phase AC voltage supply using the power cord provided. The generator is fitted with an internal transformer that can accept any supply from 100 to 240 volts AC. If the appropriate power cord is not supplied; a new plug, rated to at least 5 amps, can be fitted by a qualified electrician.
  • Page 24: Product Registration

    Product Registration Before the generator will start up for the first time the user must enter a unique 4-digit PIN code. To receive your generator’s unique PIN code, please register on the Peak website www. peakscientific.com/activate or download the Peak mobile app from the Google Play Store or Apple App Store.
  • Page 25: Home Screen

    Home Screen This is the Main Home page screen at first startup of the system. Error Icon in the top right of the screen is Normal, as pressure will be low and O2 content will be high. Touching the red triangle icon, will take you to the error page.
  • Page 26: Settings Screen

    Settings Screen This is the Settings Screen where displayed language can be selected. Service Screen This is the Service Lock Screen – only trained and authorised engineers will have the ability to enter a PIN code and access generator settings and param- eters.
  • Page 27: Screen Warning Icons

    Screen Warning Icons The Red warning triangle in the top right corner of the screen, indicates a fault condition is active on the system. Pressing the icon will take you to the Error page which will include a full description of the fault.
  • Page 28: Commissioning & Safe Start-Up Sequence

    Commissioning & Safe Start-Up Sequence Note: The side panels of the generator should not be removed during operation unless you have received training and are technically competent to manage the potential risks present. Located inside the cabinet are the vent silencers and safety relief valve, which periodically release gas at WARNING pressure and could cause injury.
  • Page 29 Once the desired tank pressure has been achieved, the system will automatically exit the “Shortcycle” mode and begin “Normal Operation”. The system will continue to produce Gas until the Standby pressure is reached, at which point the generator will enter the “Standby” mode. Once in “Standby”...
  • Page 30: Normal Operation

    Normal Operation i-Flow 7000 series gas generators are designed specifically to minimize operator involvement. As long as the system is installed as described in earlier sections and is serviced in accordance with the specified maintenance recommendations (see Service Requirements), then it should simply be a matter of turning the Generator on. Note: the generator will only produce nitrogen gas based on demand, so should typically be left in a powered state with live compressed air supply.
  • Page 31: Safe Isolation Process

    Safe Isolation Process With reference to the diagram in the Recommended Piping Layout section. To shut the system down, first close the Tank/Application isolation Valve (7) and the Generator outlet isolation Valve (5), this will isolate and preserve the gas within the Nitrogen receiver/storage/buffer tank.
  • Page 32: Service Requirements

    Service Requirements Due to the simplicity of the design and the small number of moving parts the i-Flow N2 Nitrogen Generators will have a long and trouble free life. However as with all technical equipment it should be regularly inspected and serviced as below. Service Schedule Quantity.
  • Page 33: Service Indication

    Service Indication The generator will notify the user of the service interval for the filter elements. The generator has the following Service Indication Stages:- Stage 1  When less than 30 days are remaining on the service counter the Service Due icon will show in the top right corner of the Home Screen.
  • Page 34: Peak Protected

    Peak Protected With Peak Gas Generation you invest in not only a product but peace of mind. With a network of certified Peak engineers stationed throughout the globe, Peak’s rapid response team are never far away and our commitment is to keep your generator running day in, day out, protecting your productivity.
  • Page 35: Cleaning

    Generator or material contained within it. If there is any doubt about the compatibility of decontamination or cleaning agents please contact your Peak Gas Generation representative. Cleaning should only be undertaken with the power switched off and the power cord removed from the rear of the generator.
  • Page 36: Error Messages

    Error Messages In the event of a fault condition on the system the RED warning triangle will be displayed in the top right corner of the Home Screen, pressing the Icon will take you to the Error Page. Please note that upon start-up, the RED warning triangle is normal (refer to P26 of the user manual) If no faults are currently active the Error Page will be blank, possible Error Messages are: When less than 30 days remaining on service counter, this...
  • Page 37: Troubleshooting

    Troubleshooting Problem Possible Solution • Ensure power cable is plugged into the generator and that the socket is turned on • Check the fuse in the power cable plug (if The generator will not switch on and the fitted) power switch does not illuminate. •...
  • Page 38 Page 38...
  • Page 39 Page 39...
  • Page 40 Tel: +44 141 812 8100 Fax: +44 141 812 8200 For further information on any of our generator products please contact contact@peakgas.com Copyright © Peak Scientific Instruments Ltd 2023 — i-Flow 7000 Series User Manual Rev 1 RSID 4707 EN - Alt —13/01/2023...

This manual is also suitable for:

I-flow n2 miniI-flow n2 prime

Table of Contents