Natus ® Brain Monitor & Embla Dx Series ™ User and Service Manual Table of Contents Introduction ...................................... 5 Device Description ..................................5 Intended Use, Intended User and Patient Groups ........................5 Clinical Benefit ....................................5 Medical Conditions ..................................6 Residual Risks and Side Effects ..............................
Page 4
Adding the Breakout Box to a Pouch ............................. 41 Potential Equalization Conductor ............................... 41 Connection Mode ..................................42 Connecting the Natus Base Unit and Brain Monitor & Embla Dx Series Breakout Boxes ............43 Removing the Connections ................................. 43 Connecting to the Body Position Pod ............................44 Connecting to the Luer Lock Pressure Sensor ..........................
Brain Monitor & Embla Dx Series amplifiers are designed for the acquisition of EEG and Polysomnographic (PSG) data. The amplifiers design consists of a Natus Base unit and an amplifier breakout box. The Natus Base is common for all variations of the amplifier and the breakout box varies depending on the amplifier model. There are three variations of the Natus Brain Monitor &...
Isolation transformers are intended to provide power for all system components while ensuring the safety of the operators and the patients. Carts are intended to support and facilitate the use of the Brain Monitor & Embla Dx Series Amplifiers while keeping the system mobile.
User and Service Manual Natus ® Brain Monitor & Embla Dx Series ™ General and Network Security Measures Utilize best practice security measures in your clinical environment including the following: General Security Policies Prevent physical access to the system from unauthorized persons.
Operating Principle of the Amplifier The Natus Brain Monitor & Embla Dx Series amplifiers are comprised of a base unit and a breakout box. The variation of the amplifier and feature subset is defined by the model of breakout that is connected to the base unit. It is part of a system that is made up of a personal computer, a photic stimulator, an isolation transformer, video and audio equipment, networking equipment, and mechanical supports.
Mode of operation: Continuous Environmental Conditions: Normal: 10-30°C, 30-75% rH, 700-1060hPa The Brain Monitor & Embla Dx Series Amplifiers and accessories have been designed to comply with the following national and international standards. Safety Standard of Compliance and Normative References •...
Electromagnetic Environment - Guidance RF emissions Group 1 The Brain Monitor & Embla Dx Series Amplifiers use RF energy CISPR 11 only for its internal function. Therefore, their RF emissions are very low and not likely to cause any interference in nearby...
Page 12
The Brain Monitor & Embla Dx Series Amplifiers are intended for use in the electromagnetic environment specified below. The customer or the user of the Brain Monitor & Embla Dx Series Amplifiers should assure that they are used in such an environment.
Page 13
The Brain Monitor & Embla Dx Series Amplifiers are intended for use in the electromagnetic environment specified below. The customer or the user of the Brain Monitor & Embla Dx Series Amplifiers should assure that they are used in such an environment.
User and Service Manual Natus ® Brain Monitor & Embla Dx Series ™ Table 4 - Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment Test Band Service Modulation Maximum Distance IMMUNITY frequency Power TEST LEVEL (MHz) (MHz) (V/m) 380 –390...
Inaccurate measurements may be caused by incorrect application or use. When replacing the fuse for the Brain Monitor & Embla Dx Series Amplifiers, it must be replaced by a fuse with the same type and rating as the original fuse. Replacement fuses should be purchased from Natus directly.
Refer to the Recommended Separation Distances table in the Safety & Standards Conformity section for minimum recommended separation distances. Do not connect items which are not specified as part of the Brain Monitor & Embla Dx Series Amplifier system to the system.
ELECTRICAL SHOCK HAZARD: Do NOT service the system. Refer servicing to qualified personnel only. Do NOT use repaired components without proper testing. Do not use the MPSO with the Brain Monitor & Embla Dx Series Amplifiers system for supplying power to any equipment that is not part of the system.
Page 18
Brain Monitor & Embla Dx Series ™ The Brain Monitor & Embla Dx Series Amplifiers are NOT designed to work with defibrillators. The Brain Monitor & Embla Dx systems could be damaged when used with this device. When using the Brain Monitor & Embla Dx Series Amplifiers together with electrosurgical units (cautery devices) it is recommended to connect to the acquisition PC using the TCP/IP interface.
Make sure that any platform, table, cart, or other surface used during the operation, transport, or temporary or permanent storage of the system and its components is adequate, sturdy, and safe. Natus is not responsible for any injury or damage that may result from inadequate, poorly constructed, or unapproved transports, carts, or operating surfaces.
User and Service Manual Natus ® Brain Monitor & Embla Dx Series ™ Be sure to take the appropriate Electrostatic Discharge (ESD) precautions. Disconnect the cables before moving, cabling, or performing any set up procedures. Some semiconductor (solid state) devices can be easily damaged by static electricity. Such components are commonly called Electrostatically Sensitive Devices (ESD).
Natus ® Brain Monitor & Embla Dx Series ™ User and Service Manual Description of Symbols Symbol Title Standard Title as per Symbol Standard Reference Explanation of Symbol Referenced Standard ISO 15223-1 Medical devices An indication This product is a Symbol 5.7.7...
Page 22
User and Service Manual Natus ® Brain Monitor & Embla Dx Series ™ Symbol Title Standard Title as per Symbol Standard Reference Explanation of Symbol Referenced Standard 21 CFR 801.109(b)(1) Labeling- Device is Indicates that the Prescription cleared for the...
Page 23
Natus ® Brain Monitor & Embla Dx Series ™ User and Service Manual Symbol Title Standard Title as per Symbol Standard Reference Explanation of Symbol Referenced Standard ISO 15223-1 Medical devices Legal Indicates the Symbol 5.1.1 — Symbols to be...
Page 24
User and Service Manual Natus ® Brain Monitor & Embla Dx Series ™ Symbol Title Standard Title as per Symbol Standard Reference Explanation of Symbol Referenced Standard ISO 15223-1 Medical devices Humidity limit Indicates the range Symbol 5.3.8 — Symbols to be...
Page 25
Natus ® Brain Monitor & Embla Dx Series ™ User and Service Manual Symbol Title Standard Title as per Symbol Standard Reference Explanation of Symbol Referenced Standard IEC 60601-1 Medical electrical Follow Refer to instruction Table D.2 #10 equipment – Part Instructions for manual/Booklet.
Page 26
User and Service Manual Natus ® Brain Monitor & Embla Dx Series ™ Symbol Title Standard Title as per Symbol Standard Reference Explanation of Symbol Referenced Standard Graphical symbols Protective earth; Identifies any terminal IEC 60601-1 for use on protective which is intended for Table D.1 #6...
Page 27
Natus ® Brain Monitor & Embla Dx Series ™ User and Service Manual Symbol Title Standard Title as per Symbol Standard Reference Explanation of Symbol Referenced Standard Graphical symbols "OFF" (power) Indicates disconnection IEC 60601-1 for use on from the mains, at Table D.1 #13...
Page 28
User and Service Manual Natus ® Brain Monitor & Embla Dx Series ™ Symbol Title Standard Title as per Symbol Standard Reference Explanation of Symbol Referenced Standard Classification of Ingress Protection from IEC 60529 degrees of Protection dripping water from...
Natus ® Brain Monitor & Embla Dx Series ™ User and Service Manual Specifications: Brain Monitor & Embla Dx Series Amplifiers Specifications Brain Monitor & Embla NDx Embla SDx AC Channels 64 AC channels, 6 Sensor inputs 20 AC Channels + 6 Sensor inputs...
Page 30
User and Service Manual Natus ® Brain Monitor & Embla Dx Series ™ Specifications Brain Monitor & Embla NDx Embla SDx Atmospheric Pressure: 70 kPa–106 kPa Transport and Storage -25°C to 60°C Temperature Range (-13°F to 140°F) Transport and Storage Humidity Range 10%–95%...
Power Button. Labeled with Reserved for future use. Labeled with Can be used to connect the Natus Base Unit directly to a PC and/or Laptop. Labeled with Can be used to connect the Natus Base Unit to a system network. Labeled with Ground equipotential terminal.
Natus ® Brain Monitor & Embla Dx Series ™ User and Service Manual Breakout Boxes Brain Monitor Nonin Oximeter Connection. Labeled with Base to Breakout Main Cable Connection. Labeled with Luer Lock Connection. Labeled with Body Position Pod Connection. Labeled with Patient Event Button Connection.
User and Service Manual Natus ® Brain Monitor & Embla Dx Series ™ Embla NDx Nonin Oximeter Connection. Labeled with Base to Breakout Main Cable Connection. Labeled with Luer Lock Connection. Labeled with Body Position Pod Connection. Labeled with Patient Event Button Connection. Labeled with .
Natus ® Brain Monitor & Embla Dx Series ™ User and Service Manual Embla SDx Nonin Oximeter Connection. Labeled with Base to Breakout Main Cable Connection. Labeled with Luer Lock Connection. Labeled with Body Position Pod Connection. Labeled with Patient Event Button Connection. Labeled with .
User and Service Manual Natus ® Brain Monitor & Embla Dx Series ™ Brain Monitor SpO2 Sensor Cable Clip Natus Brain Monitor breakout boxes come with a removeable wire clip connector retainer that holds the oximeter cable in place. P/N 036850...
Connect the power cable, base to breakout cable, patient event button, and any additional accessories (XactTrace belts, cannula, thermistor, snoring sensor, pulse oximeter probe, DC inputs cables etc.) as required. NOTE: Setup and Installation of the Brain Monitor & Embla Dx Series Amplifiers System should be performed by Natus qualified personnel only.
® Brain Monitor & Embla Dx Series ™ Setting the IP Address Once the Natus base unit has been powered up, press and hold power button for 3 seconds. The following user interface appears. Press the Settings button on the touchscreen to access the IP Configuration.
Touchscreen Icons When properly installed and connected to the PC, the Brain Monitor & Embla Dx Series Amplifiers displays several icons at the bottom of the touchscreen which indicate the connectivity status of the breakout boxes, the headbox, and the base units. The following table shows the different connection status indicators.
Page 40
With the Cycle feature set to ON in the NeuroWorks / SleepWorks User interface, the system will automatically measure and display impedances for each channel. Displays the information about the Breakout connected. Sends the selected note to the NeuroWorks / SleepWorks Software from the Natus Base unit.
Beginning a study Once the equipment has been installed by your Natus qualified representative and a patient has been connected to the system, a new EEG or Sleep study can be started. For details on beginning a new EEG or Sleep study, consult the NeuroWorks / SleepWorks reference manuals directly.
Potential Equalization Conductor The Natus base unit provides a potential equalization conductor for optional use. To install, connect a bus cable from the potential equalization conductor to the potential equalization busbar of the electrical insulation in the room where the system...
The following section describes how the Brain Monitor & Embla Dx Series Amplifiers system is connected. When the Natus base unit is physically connected to the breakout box using the patient cable, the Brain Monitor & Embla Dx Series amplifiers are referred to as being in Physical Connection Mode. In this mode, the data is transmitted over the wired connection to the base unit, which then transmits the data to the computer.
To Connect the Natus base unit to the Amplifier: Locate the Patient Cable from Amplifier to Base (p/n 013348) which was included with your purchase and connect it to the Natus Base unit at the connection labeled with (#1 on the image above).
Connecting to the Luer Lock Pressure Sensor The Brain Monitor and Embla Dx Series amplifiers each include a stainless steel luer lock located at the bottom end of each breakout. This luer lock allows the connection of a cannula to allow recording of airflow via pressure sensor.
Refer to the corresponding Instructions for Use for all system components prior to use. This should include, but is not limited to cameras, computers, stimulators, and software. NOTE: Transportation System setup and installation should be performed by Natus qualified personnel only. NOTE: Trolley and cart availability varies by region.
User and Service Manual Breakout Box Holster with Roll Cart Natus ErgoJust Cart Specifications For specifications and details on the ErgoJust cart, please refer to the Natus ErgoJust Installation & Functionality Guide (p/n: 019667). Maintenance Regularly inspect the trolley to ensure that casters, bolts, equipment mounting, and shelf fasteners are always secured tight.
Transportation System Specifications and Maintenance Warnings and Cautions WARNING: Only use Natus approved equipment on the trolley/cart. Non-approved equipment may compromise the function and safety of the system. WARNING: Do not tilt the trolley/cart more than 10° incline as this will compromise the stability of the trolley/cart.
NOTE: Refer to the Instructions for Use for any pulse oximeter sensors prior to use. The following topic lists the specifications for the Nonin Oximeter sensor that is used with the Brain Monitor & Embla Dx Series Amplifiers. The Brain Monitor & Embla Dx Series Amplifiers provides power and isolates the oximeter from the main power and ground.
Electrodes and accessories The Brain Monitor & Embla Dx Series Amplifiers and their components should not be immersed in water or any other fluid. To clean, use a damp cloth or cleaning/disinfecting products such as isopropyl alcohol, PDI “Sani Wipes AF3”, PDI “Sani-Cloth Plus”, Metrex CaviWipes, or AIC Wedge Wipes to wipe all surfaces, cables and accessories.
Equipment) Directive 2012/19/EU. These regulations state that electrical and electronic waste must be separately collected for the proper treatment and recovery to ensure that WEEE is reused or recycled safely. In line with that commitment Natus Medical Incorporated may pass along the obligation for take back and recycling to the end user unless other arrangements have been made.
Natus ® Brain Monitor & Embla Dx Series ™ User and Service Manual Troubleshooting If the acquired waveforms are flat, do not appear, or do not appear correctly (or as expected), try shutting down the computer for at least 10 seconds, and then set up the test again from the beginning. Shutting down and starting over resets the headbox and sometimes solves the problem.
Natus Brain Monitor Breakout 021919 Natus Embla NDx Breakout 021920 Natus Embla SDx Breakout 013348 Natus Quantum Breakout to Base Cable, 33ft (10m) 013414 Natus Quantum Breakout to Base Cable, 16ft (5m) 015170 Replacement Fuse, Natus Base Unit 010893 Medical Network Isolator with 10in (25cm) Patch Cable...
Page 55
Breakout Box Holster, Arm & Roll Stand 036850 Wire Clip, Connector Retainer EEG/PSG accessories which can be used with the Brain Monitor & Embla Dx Series Amplifiers are available for you to browse in the Natus Neurology Accessories Catalog online at www.natus.com...
Brain Monitor & Embla Dx Series ™ Getting Help Natus is committed to providing support to operate the Brain Monitor & Embla Dx Series Amplifier. Follow these steps to find a solution: Step 1: Document the Incident Carefully document the incident. If possible, note error messages, dialog box names and what you did before the problem occurred.
Natus ® Brain Monitor & Embla Dx Series ™ User and Service Manual Appendix A Digital Trigger Input Port Wiring Diagram Pin# Definition Note TRG1 Digital Trigger in (0~5V) TRG2 ‘’ TRG3 ‘’ TRG4 ‘’ TRG5 ‘’ TRG6 ‘’ TRG7 ‘’...
User and Service Manual Natus ® Brain Monitor & Embla Dx Series ™ Appendix B Headcap Connector Wiring To connect a pre-wired headcap or assembly to the headcap connector a 25-pin female type latch connector is required. This should come from the manufacturer of the pre-wired headset or assembly or a 3rd party vendor. The following shows the layout of the bulk connector pin box inputs.
Natus ® Brain Monitor & Embla Dx Series ™ User and Service Manual Sample Rate: 4096Hz...
Page 62
A Total Service Solution Standing behind every XLTEK product is Natus Medical Incorporated, an internationally respected innovator of medical products and services. Our Neurology systems are backed up by an in-house support team staffed with technical and clinical experts, 24/7 support, remote support via WebEx or VPN, the largest clinical and technical field support network in Neuro/Sleep and customized service contracts that include preventative maintenance visits and computer upgrades.