Avtech AVOZ-D7-B Instructions Manual

1000 volt, 100 amp pulse generator with ieee 488.2 / rs-232 / ethernet control

Advertisement

Quick Links

A V T E C H
E L E C T R O S Y S T E M S
L T D .
N A N O S E C O N D
W A V E F O R M
E L E C T R O N I C S
S I N C E
1 9 7 5
info@avtechpulse.com
Tel: 888-670-8729 (USA & Canada)
BOX 5120, LCD MERIVALE
http://www.avtechpulse.com/
or +1-613-686-6675 (Worldwide)
OTTAWA, CANADA K2C3H5
INSTRUCTIONS
MODEL AVOZ-D7-B
1000 VOLT, 100 AMP
PULSE GENERATOR
WITH IEEE 488.2 / RS-232 / ETHERNET CONTROL
SERIAL NUMBER:
14230

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AVOZ-D7-B and is the answer not in the manual?

Questions and answers

Summary of Contents for Avtech AVOZ-D7-B

  • Page 1 1 9 7 5 info@avtechpulse.com Tel: 888-670-8729 (USA & Canada) BOX 5120, LCD MERIVALE http://www.avtechpulse.com/ or +1-613-686-6675 (Worldwide) OTTAWA, CANADA K2C3H5 INSTRUCTIONS MODEL AVOZ-D7-B 1000 VOLT, 100 AMP PULSE GENERATOR WITH IEEE 488.2 / RS-232 / ETHERNET CONTROL SERIAL NUMBER: 14230...
  • Page 2: Warranty

    Avtech product is found to be defective, Avtech shall at its option repair or replace said defective item. This warranty does not apply to units which have been dissembled, modified or subjected to conditions exceeding the applicable specifications or ratings.
  • Page 3: Table Of Contents

    TABLE OF CONTENTS WARRANTY........................2 TECHNICAL SUPPORT....................2 TABLE OF CONTENTS....................3 INTRODUCTION.......................5 AVAILABLE OPTIONS.....................6 HIGH-VOLTAGE PRECAUTIONS..................7 SPECIFICATIONS......................8 REGULATORY NOTES....................9 FCC PART 18.......................... 9 EC DECLARATION OF CONFORMITY..................9 DIRECTIVE 2011/65/EU (RoHS)...................10 DIRECTIVE 2002/96/EC (WEEE)..................10 FIRMWARE LICENSING.......................11 INSTALLATION......................12 VISUAL CHECK........................12 POWER RATINGS........................ 12 CONNECTION TO THE POWER SUPPLY................12 PROTECTION FROM ELECTRIC SHOCK................13 ENVIRONMENTAL CONDITIONS..................14...
  • Page 4 PCB 197K - HIGH VOLTAGE DISCHARGE CIRCUIT............37 PCB 183A-S AND 183A-P CAPACITOR BANKS..............38 PCB 104G - KEYPAD / DISPLAY BOARD................39 MAIN WIRING - POSITIVE (-P) UNITS.................40 PERFORMANCE CHECK SHEET.................41 Manual Reference: /fileserver2/officefiles/instructword/avoz/AVOZ-D7-B,ed3.odt. Last modified April 12, 2022. Copyright © 2022 Avtech Electrosystems Ltd, All Rights Reserved.
  • Page 5: Introduction

    If the five outputs are connected to a common 10 Ohm load, the AVOZ-D7-B can deliver up to 100 Amps of pulsed current to the load. An output module is optionally available to simplify the connection of a single load to multiple outputs.
  • Page 6: Available Options

    AVAILABLE OPTIONS -CK5 Option: Cable kit (five RG-58C/U cables, 5ft/152cm). -OM5 Option: Output module accessory, with five input connectors and one output connector. -R5 Option: Rack-mount kit.
  • Page 7: High-Voltage Precautions

    4. Keep in mind that all cables, connectors, oscilloscope probes, and loads must have an appropriate voltage rating. Do not attempt any repairs on the instrument, beyond the fuse replacement procedures described in this manual. Contact Avtech technical support (see page 2 for contact information) if the instrument requires servicing.
  • Page 8: Specifications

    This will provide lower output currents. 3) For applications where additional resistance must be added in series with the device under test, Avtech recommends connecting multiple Ohmite (www.ohmite.com) OY- series ceramic composition resistors in parallel to create a high-power, low-inductance effective resistance. These resistors can be purchased readily at http://www.digi- key.com.
  • Page 9: Regulatory Notes

    If interference is observed, check that appropriate well-shielded cabling is used on the output connectors. Contact Avtech (info@avtechpulse.com) for advice if you are unsure of the most appropriate cabling. Also, check that your load is adequately shielded. It may be necessary to enclose the load in a metal enclosure.
  • Page 10: Directive 2011/65/Eu (Rohs)

    < 1000 ppm (0.1% by mass) DIRECTIVE 2002/96/EC (WEEE) European customers who have purchased this equipment directly from Avtech will have completed a “WEEE Responsibility Agreement” form, accepting responsibility for WEEE compliance (as mandated in Directive 2002/96/EC of the European Union and local...
  • Page 11: Firmware Licensing

    Article 9 of Directive 2002/96/EC. Customers who have purchased Avtech equipment through local representatives should consult with the representative to determine who has responsibility for WEEE compliance. Normally, such responsibilities with lie with the representative, unless other arrangements (under Article 9) have been made.
  • Page 12: Installation

    INSTALLATION VISUAL CHECK After unpacking the instrument, examine to ensure that it has not been damaged in shipment. Visually inspect all connectors, knobs, liquid crystal displays (LCDs), and the handles. Confirm that a power cord, a GPIB cable, and two instrumentation manuals (this manual and the “Programming Manual for -B Instruments”) are with the instrument.
  • Page 13: Protection From Electric Shock

    5. Do not attempt any repairs on the instrument, beyond the fuse replacement procedures described in this manual. Contact Avtech technical support (see page 2 for contact information) if the instrument requires servicing. Service is to...
  • Page 14: Environmental Conditions

    6. no pollution or only dry, non-conductive pollution. LABVIEW DRIVERS A LabVIEW driver for this instrument is available for download on the Avtech web site, at http://www.avtechpulse.com/labview. A copy is also available in National Instruments' Instrument Driver Library at http://www.natinst.com/.
  • Page 15: Fuses

    FUSES This instrument contains four fuses. All are accessible from the rear-panel. Two protect the AC prime power input, and two protect the internal DC power supplies. The locations of the fuses on the rear panel are shown in the figure below: Fuses #1 and #2 Fuse #4 Fuse #3...
  • Page 16: Dc Fuse Replacement

    DC FUSE REPLACEMENT The DC fuses may be replaced by inserting the tip of a flat-head screwdriver into the fuse holder slot, and rotating the slot counter-clockwise. The fuse and its carrier will then pop out. FUSE RATINGS The following table lists the required fuses: Recommended Replacement Part Nominal Fuses...
  • Page 17: Front Panel Controls

    FRONT PANEL CONTROLS 1. POWER Switch . This is the main power switch. When turning the instrument on, there is normally a delay of 10 seconds before anything is shown on the main display, as the internal operating system boots up. If the main menu does not appear after 30 seconds, turn off the instrument and leave it off for at least 60 seconds before applying power again.
  • Page 18 detail. 5. KEYPAD . Control Name Function MOVE This moves the arrow pointer on the display. CHANGE This is used to enter the submenu, or to select the operating mode, pointed to by the arrow pointer. ×10 If one of the adjustable numeric parameters is displayed, this increases the setting by a factor of ten.
  • Page 19: Rear Panel Controls

    REAR PANEL CONTROLS GATE TRIG RS-232 GPIB 1. AC POWER INPUT . An IEC-320 C14 three-pronged recessed male socket is provided on the back panel for AC power connection to the instrument. One end of the detachable power cord that is supplied with the instrument plugs into this socket. 2.
  • Page 20 6. GPIB Connector . A standard GPIB cable can be attached to this connector to allow the instrument to be computer-controlled. See the “Programming Manual for -B Instruments” for more details on GPIB control. 7. RS-232 Connector. A standard serial cable with a 25-pin male connector can be attached to this connector to allow the instrument to be computer-controlled.
  • Page 21: General Information

    GENERAL INFORMATION BASIC PULSE CONTROL This instrument can be triggered by its own internal clock or by an external TTL trigger signal. In either case, two output channels respond to the trigger: OUT and SYNC. The OUT channel is the signal that is applied to the load. Its amplitude and pulse width are variable.
  • Page 22 > 50 ns TRIG TTL LEVELS (external input) (0V and 3V-5V) PROPAGATION DELAY (FIXED) 100 ns, FIXED SYNC OUT 3V, FIXED DELAY > 0 PULSE WIDTH AMPLITUDE, VARIABLE MAIN OUTPUT Figure C As before, if the delay is negative, the order of the SYNC and OUT pulses is reversed. The last figure illustrates the relationship between the signal when an external TTL-level trigger is used in the PW mode.
  • Page 23: Trigger Modes

    TRIGGER MODES This instrument has four trigger modes:  Internal Trigger: the instrument controls the trigger frequency, and generates the clock internally.  External Trigger: the instrument is triggered by an external TTL-level clock on the back-panel TRIG connector.  Manual Trigger: the instrument is triggered by the front-panel “SINGLE PULSE” pushbutton.
  • Page 24: The Load And Output Module

    THE LOAD AND OUTPUT MODULE The AVOZ-D7-B has five identical outputs. These outputs can either be combined to drive a single low-impedance (10 Ohms) load, or can be used separately to drive several 50 Ohm loads simultaneously. If the -OM5 option has been ordered, an output module is provided to simplify the connection of a single load to multiple outputs.
  • Page 25: Preventing Output Stage Failure

    AVOZ-D REAL-TIME MAINFRAME SCOPE OSCILLOSCOPE PROBE OUT CONNECTOR #1 CHANNEL A OUT CONNECTOR #2 50 Ω, OUT CONNECTOR #3 OUT CONNECTOR #4 OUT CONNECTOR #5 POWER CHANNEL B To three other 50 Ω, 50 Ohm loads. TRIG SYNC INPUT OUTPUT PREVENTING OUTPUT STAGE FAILURE The output stage is protected against overload conditions by an overload circuit and fuses on the main frame back panel.
  • Page 26: Operational Check

    OPERATIONAL CHECK This section describes a sequence to confirm the basic operation of the instrument. It should be performed after receiving the instrument. It is a useful learning exercise as well. Before proceeding with this procedure, finish read this instruction manual thoroughly. Then read the “Local Control”...
  • Page 27 3. Turn on the AVOZ-D7-B. The main menu will appear on the LCD. 4. To set the AVOZ-D7-B to trigger from the internal clock at a PRF of 10 Hz:  The arrow pointer should be pointing at the frequency menu item. If it is not, press the MOVE button until it is.
  • Page 28 8. To enable the output:  Press the MOVE button until the arrow pointer is pointing at the output menu item.  Press the CHANGE button. The output submenu will appear.  Press MOVE until the arrow pointer is pointing at the “ON” choice. ...
  • Page 29: Programming Your Pulse Generator

    PROGRAMMING YOUR PULSE GENERATOR KEY PROGRAMMING COMMANDS The “Programming Manual for -B Instruments” describes in detail how to connect the pulse generator to your computer, and the programming commands themselves. A large number of commands are available; however, normally you will only need a few of these.
  • Page 30: All Programming Commands

    ALL PROGRAMMING COMMANDS For more advanced programmers, a complete list of the available commands is given below. These commands are described in detail in the “Programming Manual for -B Instruments”. (Note: this manual also includes some commands that are not implemented in this instrument.
  • Page 31 :SOURce INTernal | EXTernal | MANual | HOLD | IMMediate *CLS [no query form] *ESE <numeric value> *ESR? [query only] *IDN? [query only] *OPC *SAV 0 | 1 | 2 | 3 [no query form] *RCL 0 | 1 | 2 | 3 [no query form] *RST [no query form]...
  • Page 32: Mechanical Information

    There are no user-adjustable internal circuits. For repairs other than fuse replacement, please contact Avtech (info@avtechpulse.com) to arrange for the instrument to be returned to the factory for repair. Service is to be performed solely by qualified service personnel.
  • Page 33: Maintenance

    MAINTENANCE REGULAR MAINTENANCE This instrument does not require any regular maintenance. On occasion, one or more of the four rear-panel fuses may require replacement. All fuses can be accessed from the rear panel. See the “FUSES” section for details. CLEANING If desired, the interior of the instrument may be cleaned using compressed air to dislodge any accumulated dust.
  • Page 34: Wiring Diagrams

    WIRING DIAGRAMS WIRING OF AC POWER...
  • Page 35: Wiring Of Dc Power

    WIRING OF DC POWER...
  • Page 36: Pcb 158R6 - Low Voltage Power Supply

    PCB 158R6 - LOW VOLTAGE POWER SUPPLY...
  • Page 37: Pcb 197K - High Voltage Discharge Circuit

    PCB 197K - HIGH VOLTAGE DISCHARGE CIRCUIT...
  • Page 38: Pcb 183A-S And 183A-P Capacitor Banks

    PCB 183A-S AND 183A-P CAPACITOR BANKS 183A-S (SERIES CAPACIT OR BANK) 7 2 00 K -ND, Mfg . 7 2 0 0, 4-4 0 th read stando ff 3 /8" HV + GN D 7 2 00 K -ND, Mfg . 7 2 0 0, 4-4 0 th read stando ff 3 /8" X1 1 7 2 00 K -ND, Mfg .
  • Page 39: Pcb 104G - Keypad / Display Board

    PCB 104G - KEYPAD / DISPLAY BOARD...
  • Page 40: Main Wiring - Positive (-P) Units

    MAIN WIRING - POSITIVE (-P) UNITS...
  • Page 41: Performance Check Sheet

    PERFORMANCE CHECK SHEET...