Page 1
MANUAL Brief Operating Instructions Proline 300 HART Proline 500 – Digital Modbus RS485 Transmitter with Coriolis Sensor Transmitter with Coriolis sensor Brief Operating Instructions Manual Brief Operating Instructions Proline 500 KA01369O/06/EN/02.19 71454947 Modbus RS485 Transmitter with Coriolis sensor These instructions are Brief Operating Instructions; they are These instructions are Brief Operating Instructions;...
Page 2
Proline 500 – digital Modbus RS485 Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555...
Page 3
Proline 500 – digital Modbus RS485 Brief Operating Instructions for the device Brief Operating Instructions for the device The device consists of a transmitter and a sensor. The process of commissioning these two components is described in two separate manuals: •...
Proline 500 – digital Modbus RS485 Document information Document information Symbols used 1.1.1 Safety symbols Symbol Meaning DANGER! DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury. WARNING! This symbol alerts you to a dangerous situation.
Page 6
Document information Proline 500 – digital Modbus RS485 Symbol Meaning Protective Earth (PE) A terminal which must be connected to ground prior to establishing any other connections. The ground terminals are situated inside and outside the device: • Inner ground terminal: Connects the protectiv earth to the mains supply.
Proline 500 – digital Modbus RS485 Basic safety instructions Basic safety instructions Requirements for the personnel The personnel must fulfill the following requirements for its tasks: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task.
Basic safety instructions Proline 500 – digital Modbus RS485 NOTICE Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
Proline 500 – digital Modbus RS485 Basic safety instructions Product safety This measuring device is designed in accordance with good engineering practice to meet state- of-the-art safety requirements, has been tested, and left the factory in a condition in which it is safe to operate.
Product description Proline 500 – digital Modbus RS485 Product description The measuring system consists of a Proline 500 - digital transmitter and a Proline Promass or Cubemass Coriolis sensor. The transmitter and sensor are mounted in physically separate locations. They are interconnected by a connecting cable.
Proline 500 – digital Modbus RS485 Installation Installation For detailed information about mounting the sensor, see the Sensor Brief Operating Instructions → 3 CAUTION Ambient temperature too high! Danger of electronics overheating and housing deformation. ‣ Do not exceed the permitted maximum ambient temperature .
Installation Proline 500 – digital Modbus RS485 Wall mounting 17 (0.67) 5.8 (0.23) 5.8 (0.23) 149 (5.85) A0029054 2 Engineering unit mm (in) Depends on order code for "Transmitter housing" Order code for "Transmitter housing" • Option A, aluminum coated: L =14 mm (0.55 in) •...
Proline 500 – digital Modbus RS485 Electrical connection Electrical connection NOTICE The measuring device does not have an internal circuit breaker. ‣ For this reason, assign the measuring device a switch or power-circuit breaker so that the power supply line can be easily disconnected from the mains.
Page 14
Electrical connection Proline 500 – digital Modbus RS485 Signal cable Modbus RS485 The EIA/TIA-485 standard specifies two types of cable (A and B) for the bus line which can be used for every transmission rate. Cable type A is recommended.
Page 15
Proline 500 – digital Modbus RS485 Electrical connection 5.1.3 Connecting cable Non-hazardous area, Ex Zone 2, Class I, Division 2 Standard cable A standard cable can be used as the connecting cable. Standard cable 4 cores (2 pairs); pair-stranded with common shield Shielding Tin-plated copper-braid, optical cover ≥...
Page 16
Electrical connection Proline 500 – digital Modbus RS485 Cross-section Cable length Termination 2 x 2 x 0.50 mm 50 m (165 ft) BN WT YE GN (AWG 22) • +, – = 0.5 mm • A, B = 0.5 mm 3 x 2 x 0.50 mm...
Electrical connection Proline 500 – digital Modbus RS485 Connecting the measuring device NOTICE Limitation of electrical safety due to incorrect connection! ‣ Have electrical connection work carried out by appropriately trained specialists only. ‣ Observe applicable federal/national installation codes and regulations.
Page 19
Proline 500 – digital Modbus RS485 Electrical connection Connecting the connecting cable to the sensor connection housing Connection via terminals with order code for "Housing" Available for sensor Option A "Aluminum, coated" → 20 • Promass A, E, F, H, I, O, P, Q, S •...
Page 20
Electrical connection Proline 500 – digital Modbus RS485 Connecting the sensor connection housing via terminals 10 (0.4) 3 mm TX 20 TX 20 22 mm 24 mm A0029616 Loosen the securing clamp of the housing cover. Unscrew the housing cover.
Page 21
Proline 500 – digital Modbus RS485 Electrical connection Connecting the sensor connection housing via terminals 10 (0.4) 8 mm 22 mm 24 mm A0029613 Release the securing screw of the housing cover. Open the housing cover. Push the cable through the cable entry . To ensure tight sealing, do not remove the sealing ring from the cable entry.
Page 22
Electrical connection Proline 500 – digital Modbus RS485 Connecting the sensor connection housing via the connector A0029615 Connect the protective ground. Connect the connector.
Page 23
Proline 500 – digital Modbus RS485 Electrical connection Connecting the connecting cable to the transmitter TX 20 TX 20 10 (0.4) 22 mm 24 mm A0029597 Loosen the 4 fixing screws on the housing cover. Open the housing cover. Fold open the terminal cover.
Page 24
Electrical connection Proline 500 – digital Modbus RS485 5.2.2 Connecting the signal cable and the supply voltage cable A0028200 Terminal connection for supply voltage Terminal connection for signal transmission, input/output Terminal connection for signal transmission, input/output Terminal connection for connecting cable between sensor and transmitter Terminal connection for signal transmission, input/output;...
Proline 500 – digital Modbus RS485 Electrical connection Push the cable through the cable entry . To ensure tight sealing, do not remove the sealing ring from the cable entry. Strip the cable and cable ends. In the case of stranded cables, fit ferrules.
Electrical connection Proline 500 – digital Modbus RS485 To guarantee IP66/67 degree of protection, Type 4X enclosure, carry out the following steps after the electrical connection: Check that the housing seals are clean and fitted correctly. Dry, clean or replace the seals if necessary.
Proline 500 – digital Modbus RS485 Operation options Operation options Overview of operation options A0030213 Local operation via display module Computer with Web browser (e.g. Internet Explorer) or with operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) Mobile handheld terminal with Wireless Field Device Configurator App...
Operation options Proline 500 – digital Modbus RS485 Structure and function of the operating menu 6.2.1 Structure of the operating menu Operating menu for operators and maintenances Language Operation task-oriented Setup Diagnostics Operating menu for experts Expert function-oriented A0014058-EN 3 Schematic structure of the operating menu 6.2.2...
Proline 500 – digital Modbus RS485 Operation options Access to the operating menu via the local display X X X X X X 19.184 mA 12.5 X X X X X X Language à English 20.50 Deutsch Español Français Xxxxxxx...
Page 30
Operation options Proline 500 – digital Modbus RS485 6.3.1 Operational display Explanatory symbols for the measured value Status area • Depends on the device version, e.g.: The following symbols appear in the status area of the • : Volume flow operational display at the top right: •...
Page 31
Proline 500 – digital Modbus RS485 Operation options Numeric editor Confirms selection. Moves the input position one position to the left. Exits the input without applying the changes. Inserts decimal separator at the input position. Inserts minus sign at the input position.
System integration Proline 500 – digital Modbus RS485 Keys and meaning Reduces the contrast (brighter setting). Plus/Enter key combination (press and hold down the keys simultaneously) Increases the contrast (darker setting). Minus/Plus/Enter key combination (press the keys simultaneously) For operational display:Enables or disables the keypad lock.
Proline 500 – digital Modbus RS485 Commissioning Commissioning Function check Before commissioning the measuring device: ‣ Make sure that the post-installation and post-connection checks have been performed. • "Post-installation check" checklist → 12 • "Post-connection check" checklist → 26...
Commissioning Proline 500 – digital Modbus RS485 Configuring the measuring device The Setup menu with its submenus and guided wizards is used for fast commissioning of the device. They contain all the parameters required for configuration, such as for measurement or communication.
Proline 500 – digital Modbus RS485 Diagnostic information Diagnostic information Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display. The message about remedial measures can be called up from the diagnostic message, and contains important information on the fault.
Page 36
All information and technical specifications in this documentation have been carefully checked and compiled by the author. However, we can- not completely exclude the possibility of errors. TechnipFMC is always grateful to be informed of any errors. Contact us on the website.
Need help?
Do you have a question about the Proline 500 and is the answer not in the manual?
Questions and answers