Emc Compliance - OEM PCA-10 User Manual

Capacitor charging module
Table of Contents

Advertisement

EMC compliance

IEC 60601-1-2 Edition 4.0 (2014).
Environment of intended uses: Professional Healthcare Facility Environment
Declared EMC compliance (to be confirmed with test reports for every particular
part number):
Test
Emission (IEC 60601-1-2 section 7.1-7.2)
Radiated emission
Freq. range: 30 - 1000 MHz
Conducted emission
Freq. range: 150 kHz - 30 MHz
Harmonic current emission
Voltage changes, Voltage
fluctuations and Flicker test
Immunity (IEC 60601-1-2 section 8.9 & 8.10)
Immunity from Electrostatic
discharge (ESD)
Immunity from radiated
electromagnetic fields
Immunity from
Proximity field from wireless
communications equipment
Immunity from Electrical Fast
transient (EFT)
Immunity from Surge
Immunity from conducted
disturbances induced by RF fields
Immunity from Voltage dips,
short interruptions and voltage
variations
Note. The emission characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which
CISPR 11 class B is normally required) this equipment might not offer adequate protection to
Warning. Use of cables other than those specified or provided by the manufacturer of the
equipment may result in increased electromagnetic emission and/or decreased electromagnetic
immunity of the equipment and, correspondingly, in improper operation.
Standard
CISPR 11
IEC 61000-3-2
IEC 61000-3-3
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
radio-frequency communication services.
Class/ Severity level
Group 1 Class A
Group 1 Class A
120 VAC / 230 VAC
230 VAC
230 VAC
8 kV contact discharges &
15 kV air discharges
3.0 V/m
80 MHz  2.7 GHz, 80% AM, 1kHz
List of frequencies (Table 9),
from 9 V/m up to 28 V/m,
PM (18 Hz or 217 Hz), FM 1 kHz
± 2 kV on 230 VAC
Tr/Th – 5/50 ns, 100 kHz
±1.0 kV DM/ 2.0kV CM
on 230 VAC
Tr/Th – 1.2/50 (8/20) s
3.0 & 6.0 V
on
RMS
230 VAC
0.15÷ 80 MHz, 80% AM, 1 kHz
230 & 120 VAC mains;
0 % - 10 ms; 70% - 500 ms;
0% - 20 ms; 0% - 5sec
Test result
Complies
Complies
Complies
Complies
Complies
Complies
Complies
Complies
Complies
Complies
Complies
13