Specifications - Water Analytics AquaMetrix AM-ODO-TX Installation And Operation Manual

Optical dissolved oxygen probe
Table of Contents

Advertisement

4.2. Specifications

Parameters
Data Transmission
Range, DO
Accuracy, DO
Resolution, DO
Response Time, Cap
Units, DO
Range, Temp.
Accuracy, Temp
Resolution, Temp
Units, Temp
Salinity Compensation
Barometric Compensation
Methods
Calibration
Calibration Time
Calibration Freq
Wetted Material
Reading Rate
Power Requirements
Power Consumption
IP Rating
Compliance
1
Cycoloy is a registered trademark of SABIC GLOBAL Technologies B.V.;
2
Ryton is a registered trademark of Solvay SA.
ODO-TX Operating Manual
Probe Parameters
Optical Dissolved Oxygen Sensor
Analog (4-20mA), RS485, Modbus
0-60 mg/L; 0-600% Saturation
+/- 0.1 mg/L (0-20 mg/L)
+/-2% (20-60 mg/L)
0.01 mg/L
T63<5s, T90<45s, T95<60s
mg/L, ppm, % saturation
-5°C to 50°C (23°F to 122°F)
+/- 0.1°C
0.01°C
Celsius, Fahrenheit
Fixed or real-time capable
Fixed or real-time capable
EPA-approved methods 1002-8-2009,
1003-8-2009, 1004-8-2009 Standard
Methods 4500-O
1-2 points via RS485 Connection OR at
local controller
~1 min (for each point)
1-2 years
Physical
PVC Body, Cycoloy® (PC/ABS)
Ryton® (PPS), PC/PMMA
1 second
8 to 36 VDC
Maximum (measurement): 50 mA at
12 VDC
Idle (communication only): 2 mA at 12
VDC
IP-67 with sensor cap off; IP-68 with
sensor cap installed
EMC 2014/30/EU
IEC 61000-6-2:2005
EN 55011:2009
5
1
2
N116-155 rev 1.0