Water Analytics AquaMetrix AM-FOG Installation And Operation Manual

Portable fog analyzer

Advertisement

Quick Links

AM-FOG
Portable FOG Analyzer
N116-152 AM-FOG
Version 0.9 / 9-15-21

Advertisement

Table of Contents
loading

Summary of Contents for Water Analytics AquaMetrix AM-FOG

  • Page 1 AM-FOG Portable FOG Analyzer N116-152 AM-FOG Version 0.9 / 9-15-21...
  • Page 2: Table Of Contents

    Installation and Operation Manual Contents Introduction ............................3 Specifications ............................3 Principle of Operation ..........................4 What Comes with the AM-FOG Probe ....................4 Operating the AM-FOG Probe ........................ 4 Downloading the AM-FOG Connect App ..................4 Getting Going ..........................5 Configuring the Probe ........................
  • Page 3: Introduction

    1 Introduction The AM-FOG analyzer is the first probe to replace core samplers (e.g. Sludge Judge, Dipstick Pro) with a digital version that is portable, robust, customizable and inexpensive. It measures the level of fats, oils, and grease (FOG) in grease interceptors quickly, accurately and as cleanly as possible. The probe is the result of five years of intense product development.
  • Page 4: Principle Of Operation

    3 Principle of Operation As stated in the introduction, the AM-FOG analyzer works by measuring the capacitance of the surrounding medium. The capacitance is related to an electrical property called the dielectric constant. Materials that hold a large charge, such as air, have a high dielectric constant. Those that can’t, such as water, have a low dielectric constant.
  • Page 5: Getting Going

    For an Android phone go to the Play Store and look for AM-FOG. For an iPhone go to the Store. The app does NOT connect to the internet. Therefore, it collects NO information from you to send to us so you can rest assured that there is no downside to loading it onto your device. Getting Going 1.
  • Page 6: Configuring The Probe

    Configuring the Probe Successful connection with the probe automatically brings up the Configuration screen. There are four functions in the app: Measure, Diagnostics, Calibrate and Configuration. They are accessed by clicking on the menu icon in the upper left corner. The menu icon appears for the first time when the Configuration screen appears.
  • Page 7: Taking A Measurement

    To calibrate the probe, press the menu icon in the top left and bring up the menu screen (right). Select Calibrate. Follow the instructions: 1. Immerse the probe in a bath of water up to the top of the PVC pipe. (Do NOT immerse the head unit.) We use a 3”...
  • Page 8: Diagnostics

    Diagnostics For routine measurements the Measure screen is sufficient. To view the actual measurements that go into the Measure screen the Diagnostics screen is invaluable, especially for troubleshooting and for setting values in the Configuration screen. With the measure screen open, click on the little waveform icon on the upper right of the Measure screen or select...